Deep learning assisted measurement of echocardiographic left heart parameters: improvement in interobserver variability and workflow efficiency
Authors:
- 23 October 2023
- journal article
- research article
- Published by Springer Nature in The International Journal of Cardiovascular Imaging
- Vol. 39 (12) , 2507-2516
Abstract
Machine learning techniques designed to recognize views and perform measurements are increasingly used to address the need for automation of the interpretation of echocardiographic images. The current study was designed to determine whether a recently developed and validated deep learning (DL) algorithm for automated measurements of echocardiographic parameters of left heart chamber size and function can improve the reproducibility and shorten the analysis time, compared to the conventional methodology. The DL algorithm trained to identify standard views and provide automated measurements of 20 standard parameters, was applied to images obtained in 12 randomly selected echocardiographic studies. The resultant measurements were reviewed and revised as necessary by 10 independent expert readers. The same readers also performed conventional manual measurements, which were averaged and used as the reference standard for the DL-assisted approach with and without the manual revisions. Inter-reader variability was quantified using coefficients of variation, which together with analysis times, were compared between the conventional reads and the DL-assisted approach. The fully automated DL measurements showed good agreement with the reference technique: Bland–Altman biases 0–14% of the measured values. Manual revisions resulted in only minor improvement in accuracy: biases 0–11%. This DL-assisted approach resulted in a 43% decrease in analysis time and less inter-reader variability than the conventional methodology: 2–3 times smaller coefficients of variation. In conclusion, DL-assisted approach to analysis of echocardiographic images can provide accurate left heart measurements with the added benefits of improved reproducibility and time savings, compared to conventional methodology.Keywords
This publication has 17 references indexed in Scilit:
- Fully Automated Versus Standard Tracking of Left Ventricular Ejection Fraction and Longitudinal StrainJACC, 2015
- Recommendations for Cardiac Chamber Quantification by Echocardiography in Adults: An Update from the American Society of Echocardiography and the European Association of Cardiovascular ImagingJournal of the American Society of Echocardiography, 2015
- Variability of M-Mode Versus Two-Dimensional Echocardiography Measurements in Children With Dilated CardiomyopathyPediatric Cardiology, 2013
- The Ventricular Volume Variability Study of the Pediatric Heart Network: Study Design and Impact of Beat Averaging and Variable Type on the Reproducibility of Echocardiographic Measurements in Children with Chronic Dilated CardiomyopathyJournal of the American Society of Echocardiography, 2012
- Analysis of variability and reproducibility of echocardiography measurements in valvular aortic valve stenosis.2011
- Variability and Resource Utilization of Bedside Three-dimensional Echocardiographic Quantitative Measurements of Left Ventricular Volume in Congenital Heart DiseaseCongenital Heart Disease, 2006
- Contrast echocardiography improves the accuracy and reproducibility of left ventricular remodeling measurementsJACC, 2001
- Accuracy and reproducibility of biplane two-dimensional echocardiographic measurements of left ventricular dimensions and functionEuropean Heart Journal, 1997
- Interobserver Variability of Left Ventricular Measurements in a Population of Predominantly Obese Hypertensives Using Simultaneously Acquired and Displayed M‐Mode and 2‐D Cine EchocardiographyEchocardiography, 1997
- Limits of reproducibility of cross-sectional echocardiographic measurements of left ventricular muscle massInternational Journal of Cardiology, 1991