CdTe sensor configurations for robot assisted photon counting gamma camera
Authors:
- 1 July 2021
- journal article
- research article
- Published by IOP Publishing in Journal of Instrumentation
- Vol. 16 (07) , T07010
Abstract
In this work, different Cadmium Telluride (CdTe) sensor configurations are assessed for the usage in a robot assisted portable gamma camera. In the first part, four CdTe sensors, with thickness of 0.45 mm, 1 mm, 2 mm and 3 mm and pixel sizes of 55 μm and 110 μm, are investigated regarding their spectroscopic performance. The photon counting detector Timepix1 is hereby used. The 3 mm CdTe sensor shows increase in count rate up to a factor of 1.25 compared to a 2 mm CdTe sensor, 1.84 compared to a 1 mm CdTe sensor and up to 2.71 compared to a 0.45 mm CdTe sensor in the case of 137Cs. In the second part, the 3 mm CdTe sensor was implemented in a commercially available gamma camera, the iPIX. The system was integrated in the bomb disposal robot and tested in different scenarios. The integrated 3 mm CdTe detector measured 21.5 counts per second emitting from a 60Co source with an activity of 2.8 ± 0.07 Gbq in 20 meters distance in an open environment. The acquisition time was 116 seconds. The angular resolution was sufficient for the user to localize the radioactive isotope inside the test structure.Keywords
This publication has 30 references indexed in Scilit:
- Investigation of GaAs:Cr Timepix assemblies under high flux irradiationJournal of Instrumentation, 2015
- Measurements of charge sharing in small pixel CdTe detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2014
- Eiger: a single-photon counting x-ray detectorJournal of Instrumentation, 2014
- Imaging and spectroscopic performance studies of pixellated CdTe Timepix detectorJournal of Instrumentation, 2012
- Energy resolution and transport properties of CdTe-Timepix-AssembliesJournal of Instrumentation, 2011
- Pixelman: a multi-platform data acquisition and processing software package for Medipix2, Timepix and Medipix3 detectorsJournal of Instrumentation, 2011
- A Dual Modality Gamma Camera Using ${\rm LaCl}_{3}({\rm Ce})$ ScintillatorIEEE Transactions on Nuclear Science, 2009
- Investigation of charge carrier transport and charge sharing in X-ray semiconductor pixel detectors such as Medipix2Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2007
- Charge sharing in silicon pixel detectorsNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 2002
- New family of binary arrays for coded aperture imagingApplied Optics, 1989