TECHNETIUM-99M DTPA RENAL FLOW STUDIES IN GOLDBLATT HYPERTENSION

Authors:
  • JNJV NALLY
    JV NALLY
  • HCHS CLARKE
    HS CLARKE
  • JWJP WINDHAM
    JP WINDHAM
  • GGGP GRECOS
    GP GRECOS
  • MGML GROSS
    ML GROSS
  • WPWJ POTVIN
    WJ POTVIN

  • 1 January 1985
    • journal article
    • research article
    • Vol. 26  (8) , 917-924
    • PMID:3897482
Abstract
Computer-assisted dynamic renal studies were performed on a group of 14 mongrel dogs before and after the induction of unilateral renal artery stenosis. Ninety-second 99mTc diethylenetriaminepentaacetic acid ([99mTc]DTPA), 15-min [99mTc]DTPA and 30-min 131I orthoiodohippurate ([131I]hippuran) time-activity curves were analyzed and correlated with reduction of renal blood flow as measured by electromagnetic flow probe and PAH clearance techniques. Parameters of the 90-s [99mTc]DTPA curves found to be significantly different for the same kidney before and after stenosis were: upslope, curve width at 75% maximum, maximum activity value, and differential (stenotic/contralateral) maximum activity ratio. For blood flow reductions > 33%, the [99mTc]DTPA studies were judged diagnostic of unilateral renal artery stenosis in all cases, whereas the [131I]hippuran time-activity curves were indicative of stenosis in only 6 of 10 studies. In this model the computer-assisted 90-s [99mTc]DTPA renal flow study was found to be superior to conventional [131I]hippuran renography in the diagnosis of moderate-to-severe unilateral renal artery stenosis.