No added value for Computer-Assisted surgery to improve femoral component positioning and Patient Reported Outcomes in Hip Resurfacing Arthroplasty; a multi-center randomized controlled trial
Open Access
- 25 October 2019
- journal article
- research article
- Published by Springer Nature in BMC Musculoskeletal Disorders
- Vol. 20 (1) , 1-9
Abstract
Background: Computer Assisted Surgery (CAS) has proven to improve the accuracy in several orthopedic procedures. Therefore we used this technique to evaluate femoral component positioning in Hip Resurfacing Arthroplasty (HRA). The aim of this study was to evaluate imageless CAS compared to manually implanted femoral components and subsequently evaluates Patient Related Outcome Measures (PROMs). We hypothesized that the use of CAS optimizes the position of the femoral component and improves PROMs. Methods: This is a multicenter, single-blinded, randomized, controlled trial of two groups. In the CAS group guiding of the femoral component was done with imageless navigation. In the Conventional (control) group the femoral component was placed manually according to the preplanned position. The primary outcome measure consists of a maximum of 3 degrees difference between the postoperative Stem Shaft Angle (SSA) and preplanned SSA. Secondary outcome measures consist of the Hip disability and Osteoarthritis Outcome Scale (HOOS), the Harris Hip Score (HHS) and Visual Analogue Scale (VAS) pain score. Results: A total of 122 patients were randomized, 61 in the CAS group and 61 in the conventional group. There was no significant differences in accuracy of femoral implant position. The mean difference between the postoperative- and preplanned SSA was − 2.26 and − 1.75 degrees (more varus) respectively in the CAS and Conventional group. After surgery both groups show significant improvement in all PROMs compared to the baseline measurements, with no significant differences between the groups. Conclusion: Our cohort indicates no benefit for the use of CAS in accuracy of placement of the femoral component in HRA compared to manual implantation. There are no clinical differences in PROMs after 1 year follow up. This study showed no added value and no justification for the use of CAS in femoral component positioning in HRA. Trial registration: This trial is registered at ClinicalTrails.gov (https://clinicaltrials.gov/) on the 25th of October 2006: NCT00391937. Level of incidence: Level IIb, multicenter randomized controlled trial.Keywords
This publication has 38 references indexed in Scilit:
- The First SICOT Oral Presentation Award 2011: Imageless computer-assisted femoral component positioning in hip resurfacing: a prospective randomised trialInternational Orthopaedics, 2013
- Independent predictors of revision following metal-on-metal hip resurfacingThe Journal of Bone and Joint Surgery. British volume, 2012
- Insufficient Acetabular Version Increases Blood Metal Ion Levels after Metal-on-metal Hip ResurfacingClinical Orthopaedics and Related Research, 2011
- Avoiding Short-term Femoral Neck Fracture With Imageless Computer Navigation for Hip ResurfacingClinical Orthopaedics and Related Research, 2011
- Femoral Component Positioning in Hip Resurfacing With and Without NavigationClinical Orthopaedics and Related Research, 2009
- Component Alignment in Hip Resurfacing Using Computer Navigation2008
- Imageless navigation of hip resurfacing arthroplasty increases the implant accuracyInternational Orthopaedics, 2007
- Birmingham hip resurfacing arthroplastyThe Journal of Bone and Joint Surgery. British volume, 2005
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- The validation of visual analogue scales as ratio scale measures for chronic and experimental painPain, 1983