Determination of Mercury in Fish Tissue Using a Minianalyzer Based on Cold Vapor Atomic Absorption Spectrometry at the 184.9 nm Line
Open Access
- 10 September 2007
- journal article
- Published by Springer Nature in Analytical Sciences
- Vol. 23 (9) , 1121-1125
Abstract
A sensitive method was proposed and optimized for the determination of total mercury in fish tissue by using wet digestion, followed by cold vapor atomic absorption spectrometry (CVAAS) at the main resonance line of mercury (184.9 nm). The measurements were made using a new type of a non-dispersive mercury minianalyzer. This instrument was initially designed and built for atmospheric mercury-vapor detection. For determining mercury in aqueous samples, the minianalyzer was linked with a mercury/hydride system, Perkin Elmer Model MHS-10. To check the method, the analyzed samples were spiked with a standard solution of mercury. The recoveries of mercury spiked to wet fish tissue were >90% for 0.5 - 0.8 g samples. The results showed a better sensitivity (about 2.5 times higher) when using the mercury absorption line at 184.9 nm compared with the sensitivity obtained by conventional CVAAS at 253.7 nm.Keywords
This publication has 15 references indexed in Scilit:
- Determination of inorganic and total mercury by vapor generation atomic absorption spectrometry using different temperatures of the measurement cellSpectrochimica Acta Part B: Atomic Spectroscopy, 2005
- Development of a sequential injection system for trace mercury determination by cold vapour atomic absorption spectrometry utilizing an integrated gas?liquid separator/reactorTalanta, 2004
- Routine, automated determination of inorganic and total mercury in multimedia using cold vapour atomic absorption spectrometryChemosphere, 2004
- Solar blind photocell--a simple element specific detector in Hg, As and Se speciation.Analytical and Bioanalytical Chemistry, 2000
- Sequential injection analysis system for determination of mercury by cold-vapor atomic absorption spectroscopyAnalytica Chimica Acta, 1997
- Gas chromatography coupled with atomic absorption spectrometry — a sensitive instrumentation for mercury speciationSpectrochimica Acta Part B: Atomic Spectroscopy, 1996
- Novel instrumentation of a nondispersive vacuum ultraviolet atomic absorption spectrophotometer for mercurySpectrochimica Acta Part B: Atomic Spectroscopy, 1981
- Quecksilberbestimmung durch flammenlose ASS auf der Resonanzlinie 184,9 nmSpectrochimica Acta Part B: Atomic Spectroscopy, 1979
- Neues Quecksilber-Spektrometer zur Messung von Quecksilberdampf in LuftAnalytical and Bioanalytical Chemistry, 1979
- Non-flame atomic absorption in the vacuum ultraviolet regionAnalytica Chimica Acta, 1972