A Dual-Adsorbent Preconcentrator for a Portable Indoor-VOC Microsensor System
Authors:
- 6 June 2001
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 73 (14) , 3449-3457
Abstract
The development and testing of a miniature dual-adsorbent preconcentrator for a microsensor-based analytical system designed to determine complex volatile organic chemical (VOC) mixtures encountered in indoor working environments at low part-per-billion levels is described. Candidate adsorbents were screened for thermal-desorption bandwidth and breakthrough volume against 20 volatile organic vapors and subsets thereof as a function of several relevant variables. A glass capillary (1.1 mm i.d.) packed with 3.4 mg of Carbopack X and 1.2 mg of Carboxen 1000 provides sufficient capacity for a 1-L dry-air sample containing all 20 vapors at concentrations of 100 ppb as well as providing a composite half-height peak width of <3 s at a desorption temperature of 300 °C and a flow rate of 4 mL/min. Required adsorbent masses increase to 7.0 and 1.5 mg, respectively, for the same mixture at component concentrations of 1 ppm. Vapor breakthrough volumes for the Carbopack X are unaffected by humidity from 0 to 100%RH, but those for the Carboxen 1000 are significantly reduced, requiring an additional 0.9 mg of the latter to avoid premature breakthrough at the 100 ppb level. Good peak shapes, efficient chromatographic separation of preconcentrated sample mixture components, and detection limits in the low-parts-per-billion range using an integrated surface-acoustic-wave (SAW) sensor are achieved. Preconcentrator design and operating parameters are considered in terms of the vapor bed-residence times and breakthrough volumes in the context of the modified Wheeler equation.Keywords
This publication has 24 references indexed in Scilit:
- Dry purge for the removal of water from the solid sorbents used to sample volatile organic compounds from the atmospheric airThe Analyst, 2000
- Air Sampling with Porous Solid-Phase Microextraction FibersAnalytical Chemistry, 2000
- Evaluating porous-layer open-tubular capillaries as vapor preconcentrators in a microanalytical systemSensors and Actuators B: Chemical, 2000
- Peer Reviewed: Advancing the Science of Column Selectivity.Analytical Chemistry, 2000
- Analyzing organic vapors in exhaled breath using a surface acoustic wave sensor array with preconcentration: Selection and characterization of the preconcentrator adsorbentAnalytica Chimica Acta, 1998
- Total Volatile Organic Compounds (TVOC) in Indoor Air Quality Investigations*Indoor Air, 1997
- Hybrid Organic/Inorganic Copolymers with Strongly Hydrogen-Bond Acidic Properties for Acoustic Wave and Optical SensorsChemistry of Materials, 1997
- Multisorbent thermal desorption/gas chromatography/mass selective detection method for the determination of target volatile organic compounds in indoor airEnvironmental Science & Technology, 1992
- Chemical Cartridge Respirator Performance: 1,3-ButadieneAIHA Journal, 1987
- Performance of fixed-bed catalytic reactors with poison in the feedJournal of Catalysis, 1969