Beam characteristics of a retrofitted double‐focused multileaf collimator
Authors:
- Indra J. DasIDIndra J. Das
- Gregory E. DesobryGDGregory E. Desobry
- Shawn W. McNeeleySMShawn W. McNeeley
- Elizabeth C. ChengECElizabeth C. Cheng
- Timothy E. SchultheissTSTimothy E. Schultheiss
- 1 September 1998
- journal article
- Published by Wiley in Medical Physics
- Vol. 25 (9) , 1676-1684
Abstract
Multileaf collimators (MLCs) are generally believed to be convenient and cost-effective tools for intensity modulation and conformal therapy. They are becoming a standard feature on new accelerators; however, the older units can be retrofitted with modern MLCs. Before such a unit can be clinically used, the beam characteristics must be verified. In this study the beam characteristics of a Siemens double-focused MLC retrofitted to an MD2 linear accelerator are presented. The head leakage along with inter- and intra-leaf radiation transmission were measured using film. The collimator (Sc), phantom (Sp), total (Scp) scatter factors, central axis depth dose, beam profiles for off-axis ratios, penumbra, and surface dose were evaluated for square, rectangular, and irregularly shaped fields. The maximum head leakage was estimated to be < 0.05% in any plane at a distance of 1 m and maximum transmission through the MLC leaves was estimated to be < 1.4% and < 1.1% for the 10 MV and 6 MV beams, respectively. The maximum differences between pre- and post-MLC installation data for the Sc and Scp were < or = 0.7% and < or = 1.4%, respectively. Similarly, the percent depth dose data for all fields and both beam energies were within 1.5% of the original data. The beam profiles measured at various depths were also in agreement with those of the pre-MLC installation data. The measured beam penumbra (20%-80%) showed a range of 7.8 mm-11.0 mm for the 6 MV and 8.4 mm-11.1 mm for the 10 MV beams from smallest to largest fields. These ranges differ by less than a millimeter from those of the old data. The surface dose measurements were slightly lower than the conventional jaw values suggesting that MLC does not produce significant electron contamination. It is concluded that the retrofitted MLC maintains the integrity of the original beam and may provide a cost-effective conformal therapy.Keywords
This publication has 43 references indexed in Scilit:
- CT–MRI image fusion for 3D conformal prostate radiotherapy: use in patients with altered pelvic anatomyBritish Journal of Radiology, 1996
- Changes to dose in the build-up region when using multi-leaf collimators in place of lead blocks supported on an accessory trayRadiotherapy and Oncology, 1995
- Multileaf collimation versus conventional shielding blocks: a time and motion study of beam shaping in radiotherapyRadiotherapy and Oncology, 1995
- Impact of a multileaf collimator on treatment morbidity in localized carcinoma of the prostateInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1994
- The design and performance characteristics of a multileaf collimatorPhysics in Medicine & Biology, 1994
- The effect of secondary electron spread on the penumbra in high energy photon beam therapyBritish Journal of Radiology, 1990
- Benefits expected from simple conformal radiotherapy in the treatment of pelvic tumoursRadiotherapy and Oncology, 1988
- Design Principles and Clinical Possibilities with A New Generation of Radiation Therapy Equipment: A ReviewActa Oncologica, 1987
- Electron contamination from photon beam collimatorsRadiotherapy and Oncology, 1986
- Electron contamination from different materials in high energy photon beamsPhysics in Medicine & Biology, 1985