Manual Individualization of the Dialysate flow according to blood flow: Effects on the Hemodialysis dose delivered and on Dialysate consumption
Authors:
Preprint
- 8 April 2020
- preprint
- Published by Springer Nature in Research Square
Abstract
Background: The objective of this work was to assess the impact of the decrease in dialysate flow rate on the dialysis dose delivered (spKt /V) to chronic hemodialysis patients and to estimate the resulting water saving. Methods: It was a prospective 4-week-period study that included chronic hemodialysis patients with clinical and hemodynamic stability. The patients successively underwent hemodialysis with a dialysate flow rate of 500 ml / min, at 1, 1.2 and 1.5 times the blood flow rate. Each dialysate flow rate was applied for one week. During these 4 weeks, the following parameters were kept constant: duration of dialysis, blood flow rate, anticoagulation, membrane nature and surface. Results: Forty-five chronic hemodialysis patients were included with a mean age of 48.4 ± 12.07 years. The weekly average spKt/V was statistically higher with a dialysate flow rate at 1.5 times the blood flow rate compared to the dialysate flow at 500 mL / min (p = 0.001). The proportion of patients achieving a standardized dialysis dose ≥ 1.4 was statistically higher with dialysate flow at 500 mL / min (64.4%) compared to dialysate flow at 1 or 1.2 times the blood flow rate which were 57.8% and 55.6%, respectively. It was statistically higher with a dialysate flow at 1.5 times the blood flow (93.3%) compared to the dialysate flow at 500 mL / min (p = 0.036). The dialysate volume used with a dialysate flow rate of 500 mL / min was higher compared to the other dialysate flow rates (p = 0.0001). Conclusions: An adequate dialysis dose could be achieved with a dialysate flow rate of 1.5 times the blood flow rate, thereby saving significant amount of water.Keywords
All Related Versions
This publication has 13 references indexed in Scilit:
- Green nephrologyNature Reviews Nephrology, 2020
- SP491EFFECTS OF DECREASING DIALYSIS FLUID FLOW RATE ON DIALYSIS EFFICACY AND INTERDIALYTIC WEIGHT GAIN IN CHRONIC PATIENTS WITH HEMODIALYSIS - FLUGAIN StudyNephrology Dialysis Transplantation, 2018
- A Model to Predict Optimal Dialysate FlowTherapeutic Apheresis and Dialysis, 2012
- Dialysate saving by automated control of flow rates: Comparison between individualized online hemodiafiltration and standard hemodialysisHemodialysis International, 2011
- Using water wisely: New, affordable, and essential water conservation practices for facility and home hemodialysisHemodialysis International, 2009
- Effect of Dialysis Dose and Membrane Flux in Maintenance HemodialysisNew England Journal of Medicine, 2002
- Ultrapure dialysis fluid slows loss of residual renal function in new dialysis patientsNephrology Dialysis Transplantation, 2002
- Dialysate Flow Distribution in Hollow Fiber Hemodialyzers with Different Dialysate Pathway ConfigurationsThe International Journal of Artificial Organs, 2000
- Mortality and Duration of Hemodialysis TreatmentJAMA, 1991
- A mechanistic analysis of the National Cooperative Dialysis Study (NCDS)Kidney International, 1985