Abstract
Ce-doped sol-gel silica glasses were prepared for real-time X-ray dose or dose rate monitoring using radioluminescence (RL) and optically stimulated luminescence (OSL). A small piece of this glass scintillator was connected to two optical fibers, allowing a remote measurement of the ionizing radiation dose through RL and OSL signals. By using this Ce-doped material, we particularly demonstrated that the RL signal shows a linear behavior at least up to dose rates of 1.2 Gy/s. Furthermore, the linear dose response of the OSL signal can be extended up to 500 Gy, independently of the used dose rate.
Funding Information
  • Agence Nationale de la Recherche (ANR-11-LABX-0007)
  • Agence Nationale de la Recherche (ANR-11-EQPX-0017)
  • Ministry of Higher Education and Research
  • Hauts-de-France Regional Council
  • European Regional Development Fund (CPER Photonics for Society P4S)
  • IRCICA Institute
  • CERLA platform of the University of Lille

This publication has 37 references indexed in Scilit: