Numerical Modeling of Ion Transport in an ESI-MS System
- 22 March 2014
- journal article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 25 (5) , 820-831
Abstract
Gas and ion transport in the capillary-skimmer subatmospheric interface of a mass spectrometer, which is typically utilized to separate unevaporated micro-droplets from ions, was studied numerically using a two-step approach spanning multiple gas dynamic regimes. The gas flow in the heated capillary and in the interface was determined by solving numerically the Navier-Stokes equation. The capillary-to-skimmer gas/ion flow was modeled through the solution of the full Boltzmann equation with a force term. The force term, together with calculated aerodynamic drag, determined the ion motion in the gap between the capillary and skimmer. Three-dimensional modeling of the impact of the voltage applied to the Einzel lens on the transmission of doubly charged peptide ions through the skimmer orifice was compared with experimental data obtained in the companion study. Good agreement between measured and computed signals was observed. The numerical results indicate that as many as 75% of the ions that exit from the capillary are lost on the conical surface of the skimmer or capillary outer surface because of the electrostatic force and plume divergence. FigureKeywords
This publication has 40 references indexed in Scilit:
- Atmospheric Pressure Ion Source Development: Experimental Validation of Simulated Ion Trajectories within Complex Flow and Electrical FieldsJournal of the American Society for Mass Spectrometry, 2013
- Evaporation and Discharge Dynamics of Highly Charged Multicomponent Droplets Generated by Electrospray IonizationThe Journal of Physical Chemistry A, 2009
- The ion funnel: Theory, implementations, and applicationsMass Spectrometry Reviews, 2009
- Numerical Simulations of Capillary Aerosol Generation: CFD Model Development and Comparisons with Experimental DataAerosol Science and Technology, 2007
- Interacting effects of uniform flow, plane shear, and near-wall proximity on the heat and mass transfer of respiratory aerosolsInternational Journal of Heat and Mass Transfer, 2004
- Droplet Evaporation and Discharge Dynamics in Electrospray IonizationThe Journal of Physical Chemistry A, 2002
- A continuum model for ion evaporation from a drop: effect of curvature and charge on ion solvation energyAnalytica Chimica Acta, 2000
- Ionization conditions and ion formation in electrohydrodynamic mass spectrometryInternational Journal of Mass Spectrometry and Ion Processes, 1995
- Electrohydrodynamic ionization mass spectrometry: Review of instrumentation, mechanisms and applicationsJournal of Electrostatics, 1978
- Shock standoff distances and Mach-disk diameters in underexpanded sonic jetsAIAA Journal, 1971