Round Well Inset for Uniform Electric Field Distribution in Electroporation Applications
Authors:
- Praveen SahuPSPraveen Sahu
- Marco BarozziMarco Barozzi
- Paolo Di BarbaPaolo Di Barba
- Maria Evelina MognaschiMMMaria Evelina Mognaschi
- Monica La MuraMonica La Mura
- Patrizia LambertiPatrizia Lamberti
- Michele ForzanMFMichele Forzan
- Maria Teresa ConconiMCMaria Teresa Conconi
- Ignacio CamarilloICIgnacio Camarillo
- Raji SundararajanRaji Sundararajan
- Elisabetta SieniElisabetta Sieni
Open Access
- 19 February 2025
- journal article
- research article
- Published by MDPI AG in Bioengineering
- Vol. 12 (2) , 203
Abstract
Uniform electric field distribution in electroporation is crucial for achieving the effective localized delivery of drug molecules. Currently, in vitro electroporation studies on adherent cells lack a standardized test setup for obtaining consistent and repeatable results, unlike in vitro electroporation studies on cell suspensions that use electroporation cuvettes, which provide uniform electric field distribution. Considering this, we designed, built, and tested a novel inset design for low-volume round well plates, such as the 24- and 96-well plates which are most commonly used in cell culture labs. The inset design was realized using 3D printing and experimentally tested using potato phantoms and HeLa cells. Finite element analysis (FEA) was used to compute the electric field distribution in the round well plates with and without the inset. The FEA indicated that the electric field contour map at the bottom of the well with the inset had a more uniform electric field distribution, with an average value close to the expected 1000 V/cm. In contrast, it was only 840 V/cm without the inset, indicating non-uniform electric field distribution. Uniform electric fields were also obtained using the inset for the potato phantoms and the HeLa cells, indicating the merit of the inset and its usability with low-volume cell culture well plates, which enable the transfer of cells for various assays without additional steps, as well as its cost-effectiveness.Keywords
Funding Information
- EU’s Horizon Europe research and innovation program (HORIZON-MSCA-2021-SE-01)
This publication has 37 references indexed in Scilit:
- Development of 3D melanoma cultures on a hyaluronic acid-based scaffold with synthetic self-assembling peptides: Electroporation enhancementBioelectrochemistry, 2023
- Investigating the effect of electrode orientation on irreversible electroporation with experiment and simulationInternational Journal of Computer Assisted Radiology and Surgery, 2022
- Enhanced Antiproliferation Potency of Electrical Pulse-Mediated Metformin and Cisplatin Combination Therapy on MDA-MB-231 CellsApplied Biochemistry and Biotechnology, 2021
- A prospective, multi-institution assessment of irreversible electroporation for treatment of locally advanced pancreatic adenocarcinoma: initial outcomes from the AHPBA pancreatic registryHPB, 2019
- Verification of Electroporation Models Using the Potato Tuber as In Vitro SimulationJournal of Medical and Biological Engineering, 2018
- Electric field distribution in relation to cell membrane electroporation in potato tuber tissue studied by magnetic resonance techniquesInnovative Food Science & Emerging Technologies, 2016
- “Cold” electroporation in potato tissue induced by pulsed electric fieldJournal of Food Engineering, 2013
- A Three-Dimensional In Vitro Tumor Platform for Modeling Therapeutic Irreversible ElectroporationBiophysical Journal, 2012
- Magnetic Resonance Imaging Characteristics of Nonthermal Irreversible Electroporation in Vegetable TissueThe Journal of Membrane Biology, 2010
- Tissue Ablation with Irreversible ElectroporationAnnals of Biomedical Engineering, 2005