A rare case of recurrent pacemaker allergic reaction
Authors:
Open Access
- 1 January 2015
- journal article
- Published by Medknow in Heart Views
- Vol. 16 (2) , 59-61
Abstract
Allergic reactions to pacemaker device components are uncommon. However, when they occur, they usually mimic pacemaker infection, which results in multiple device replacements and increased morbidity burden. Here we present a 40-year-old female with pacemaker insertion due to complete heart block and who had multiple device replacements because of allergic sensitivity to various pacemaker component-encasing materials, confirmed by allergic testing to these materials. She had complete resolution of her symptoms after replacement with gold-plated device, to which she was not allergic.Keywords
This publication has 6 references indexed in Scilit:
- Polytetrafluoroethylene-Coated Pacemaker Leads as Surgical Management of Contact Allergy to SiliconeThe Annals of Thoracic Surgery, 2014
- Gold-coated pacemaker implantation after allergic reactions to pacemaker compoundsEP Europace, 2009
- Allergy to Pacemaker Silicone Compounds: Recognition and Surgical ManagementThe Annals of Thoracic Surgery, 2009
- Hypersensitivity reactions associated with endovascular devicesContact Dermatitis, 2008
- Pacemaker Component Allergy: Case Report and Review of the LiteratureJournal of Interventional Cardiac Electrophysiology, 2002
- Pacemaker contact sensitivity: Clinical recognition and managementThe Annals of Thoracic Surgery, 1994