Comparison of Refractive Prediction Accuracy With Three Optical Devices
Authors:
- Shi-Ming ChengSCShi-Ming Cheng
- Xi LiXLXi Li
- Jia-Sheng ZhangJZJia-Sheng Zhang
- Jian-Qi MeiJMJian-Qi Mei
- Gui-Lian ShiGSGui-Lian Shi
- Junhai LinJLJunhai Lin
- Tian-Tian LiTLTian-Tian Li
- 1 January 2023
- journal article
- research article
- Published by SLACK, Inc. in Journal of Refractive Surgery
- Vol. 39 (1) , 48-55
Abstract
Purpose: To investigate refractive prediction accuracy with the OA-2000 (Tomey), Anterion (Heidelberg Engineering), and IOLMaster 500 (Carl Zeiss Meditec AG) in patients with cataract. Methods: Patients with cataract referred for phacoemulsification were enrolled and scanned with the OA-2000, Anterion, and IOLMaster 500 in random order. The success rate of axial length (AL) measurements per device was calculated and a chi-square test was used to identify the differences in acquisition rate between the three devices. The Bland-Altman method was used to appraise the agreement of biometric parameters between the three devices. Four different formulas (Barrett Universal II [BUII], Haigis, Holladay 1, and SRK/T) were included in the study. The parameters of refractive prediction accuracy comprised predictive error (PE), absolute PE (AE), and percentages of eyes with a PE within ±0.50, ±0.75, and ±1.00 diopters (D). Results: The acquisition rates of AL measurements with the OA-2000 and Anterion were 97.35% and 94.70%, respectively (chi-square = 3.82, P > .05). A significantly lower acquisition rate of 84.82% was obtained with the IOLMaster 500 compared with the other two devices ( P < .05). Bland-Altman analysis identified good agreement between the three biometers with narrow 95% limits of agreement for flat and steep keratometry (K1 and K2), anterior chamber depth (ACD), and AL. For PE, only the differences between the Anterion and IOLMaster 500 with the Barrett UII and Haigis formulas were statistically significant ( P < .05). The three devices revealed no statistically significant differences in MAE, MedAE, and the proportion of eyes with a PE within ±0.50, ±0.75, and ±1.00 D ( P > .05). Conclusions: The OA-2000 and Anterion showed a similarly higher acquisition rate of AL measurements than the IOLMaster 500 in patients with cataract. Good agreement for K1, K2, ACD, and AL was found between the three biometers. Regarding refractive prediction accuracy, the Anterion did not significantly outperform both the OA-2000 and IOLMaster 500. [ J Refract Surg . 2023;39(1):48–55.]Keywords
This publication has 28 references indexed in Scilit:
- Axial Length Measurement Failure Rates With Biometers Using Swept-Source Optical Coherence Tomography Compared to Partial-Coherence Interferometry and Optical Low-Coherence InterferometryAmerican Journal of Ophthalmology, 2017
- Axial Length Measurement Failure Rates with the IOLMaster and Lenstar LS 900 in Eyes with CataractPLOS ONE, 2015
- Comparison of the Zeiss IOLMaster and applanation A‐scan ultrasound: biometry for intraocular lens calculationClinical & Experimental Ophthalmology, 2003
- A new non-contact optical device for ocular biometryBritish Journal of Ophthalmology, 2002
- Reproducibility of optical biometry using partial coherence interferometry: intraobserver and interobserver reliabilityJournal of Cataract & Refractive Surgery, 2001
- Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to HaigisGraefe's Archive for Clinical and Experimental Ophthalmology, 2000
- Partial coherence interferometry: a novel approach to biometry in cataract surgeryAmerican Journal of Ophthalmology, 1998
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 1992
- Development of the SRK/T intraocular lens implant power calculation formulaJournal of Cataract & Refractive Surgery, 1990
- A three-part system for refining intraocular lens power calculationsJournal of Cataract & Refractive Surgery, 1988