Surface dose variations in 6 and 10 MV flattened and flattening filter‐free (FFF) photon beams
Open Access
- 8 September 2016
- journal article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 17 (5) , 293-307
Abstract
As the use of linear accelerators operating in flattening filter-free (FFF) modes becomes more widespread, it is important to have an understanding of the surface doses delivered to patients with these beams. Flattening filter removal alters the beam quality and relative contributions of low-energy X-rays and contamination electrons in the beam. Having dosimetric data to describe the surface dose and buildup regions under a range of conditions for FFF beams is important if clinical decisions are to be made. An Elekta Synergy linac with standard MLCi head has been commissioned to run at 6 MV and 10 MV running with the flattening filter in or out. In this linac the 6 MV FFF beam has been energy-matched to the clinical beam on the central axis ( ). The 10 MV beam energy has not been adjusted. The flattening filter in both cases is replaced by a thin (2 mm) stainless steel plate. A thin window parallel plate chamber has been used to measure a comprehensive set of surface dose data in these beams for variations in field size and SSD, and for the presence of attenuators (wedge, shadow tray, and treatment couch). Surface doses are generally higher in FFF beams for small field sizes and lower for large field sizes with a crossover at at 6 MV and at 10 MV. This trend is also seen in the presence of the wedge, shadow tray, and treatment couch. Only small differences ( ) are seen between the beams on varying SSD. At both 6 and 10 MV the filter-free beams show far less variation with field size than conventional beams. By removing the flattening filter, a source of contamination electrons is exchanged for a source of low-energy photons (as these are no longer attenuated). In practice these two components almost balance out. No significant effects on surface dose are expected by the introduction of FFF delivery. PACS number(s): 87.53.Bn, 87.55.ne, 87.56.bd
Keywords
This publication has 32 references indexed in Scilit:
- A 4 MV flattening filter-free beam: commissioning and application to conformal therapy and volumetric modulated arc therapyPhysics in Medicine & Biology, 2011
- Dose perturbations by two carbon fiber treatment couches and the ability of a commercial treatment planning system to predict these effectsMedical Physics, 2009
- Dosimetric characteristics of 6 and 10MV unflattened photon beamsRadiotherapy and Oncology, 2009
- The effect of gantry angle on megavoltage photon beam attenuation by a carbon fiber couch insertMedical Physics, 2005
- Validation of Monte Carlo calculated surface doses for megavoltage photon beamsMedical Physics, 2005
- Effects of beam modifiers and immobilization devices on the dose in the build-up regionInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1994
- Design of parallel plate ion chambers for buildup measurements in megavoltage photon beamsMedical Physics, 1992
- Investigations on the near surface dose for three 10‐MV x‐ray beam accelerators with emphasis on the reduction of electron contaminationMedical Physics, 1988
- The polarity effect for commercially available plane‐parallel ionization chambersMedical Physics, 1987
- Electron contamination from photon beam collimatorsRadiotherapy and Oncology, 1986