Off‐iso Winston‐Lutz test on seven linear accelerators
Authors:
Open Access
- 23 July 2024
- journal article
- research article
- Published by Wiley in Journal of Applied Clinical Medical Physics
- Vol. 25 (10) , e14470
Abstract
Purpose: The aim of this study is to find optimal gantry, collimator, and couch angles for performing single isocenter, multiple target stereotactic radiosurgery (SIMT‐SRS). Nineteen angle sets were tested across seven linear accelerators for radiation‐isocenter coincidence and off‐isocenter coincidence. The off‐isocenter Winston‐Lutz test was performed to evaluate the accuracy of isocenter alignment for each angle set, and optimal angle sets as well as maximum off‐isocenter distance to target for each angle set was determined. The influence of simulated patient weight on off‐iso Winston‐Lutz test accuracy was also inspected.Method: The SNC MultiMet‐WL phantom and MultiMet‐WL QA Software v2.1 were used for the direct measurement and analysis of the off‐iso Winston‐Lutz test (also referred to as Winston‐Lutz‐Gao test). A two‐step method was developed to ensure precise initial placement of the target. Nineteen beams were delivered at 6X energy and 2 × 2 cm field size to each of six targets on the MultiMet Cube with couch kicks at five cardinal angles (90°, 45°, 0°, 315°, and 270°). To reduce imaging uncertainty, only EPID was used in target alignment and test image acquisition. A total of 200 Ibs (90.7 kg) of weight was also used to mimic patient weight. All tests were performed on both the free table and the weighted table.Results: For two new TrueBeam machines, the maximum offset was within the 1 mm tolerance when the off‐iso distance is less than 7 cm. Two older VitalBeam machines exhibited unfavorable gantry, couch, and collimator (GCC) angle sets: Linac No. 3 at (0,90,0), (0,270,0) and Linac No. 4 at (0,45,45) and (0,90,0). The C‐Series Linacs failed in the majority of GCC angle sets, with Linac No. 5 exhibiting a maximum offset of 1.53 mm. Four of seven machines show a clear trend that offset increases with off‐isocenter distance. Additionally, the IGRT table was less susceptible to the addition of simulated patient weight than the ExactCouch.Conclusion: Among the seven linear accelerators addressed, newer model machines such as the Varian TrueBeam were more precise than older models, especially in comparison to the C‐Series Linacs. The newer machines are more suitable for delivering SIMT‐SRS procedures in all GCC angle sets, and the results indicate that newer TrueBeams are capable of performing SIMT‐SRS procedures at all angle sets for targets of off‐iso distances up to 7 cm. The trend that offset between the target center and radiation field center increases with off‐iso distance, however, does not always hold true across machines. This may be comprised by the EPID's severe off‐axis horn effect. Lastly, the IGRT couch was less susceptible to patient weight compared to ExactCouch in the off‐isocenter Winston‐Lutz test.Keywords
This publication has 10 references indexed in Scilit:
- Experimental determination of set-up displacements in anthropomorphic phantom in single-isocentre radiosurgery for multiple brain metastases by off-axis Winston–Lutz test: ExacTracTM v.6 versus DynamicTMJournal of Radiotherapy in Practice, 2023
- AAPM Task Group Report 307: Use of EPIDs for Patient‐Specific IMRT and VMAT QAMedical Physics, 2023
- A simplified and effective off‐axis Winston–Lutz for single‐isocenter multi‐target SRSJournal of Applied Clinical Medical Physics, 2022
- A customizable, open-source Winston-Lutz system for multi-target, single isocentre radiotherapyBiomedical Physics & Engineering Express, 2022
- Detection of rotational errors in single‐isocenter multiple‐target radiosurgery: Is a routine off‐axis Winston–Lutz test necessary?Journal of Applied Clinical Medical Physics, 2022
- A novel method to determine linac mechanical isocenter position and size and examples of specific QA applicationsJournal of Applied Clinical Medical Physics, 2021
- An integrated quality assurance phantom for frameless single-isocenter multitarget stereotactic radiosurgeryPhysics in Medicine & Biology, 2020
- Development of a dedicated phantom for multi‐target single‐isocentre stereotactic radiosurgery end to end testingJournal of Applied Clinical Medical Physics, 2018
- The spatial accuracy of two frameless, linear accelerator‐based systems for single‐isocenter, multitarget cranial radiosurgeryJournal of Applied Clinical Medical Physics, 2017
- Technical Note: Evaluation of the systematic accuracy of a frameless, multiple image modality guided, linear accelerator based stereotactic radiosurgery systemMedical Physics, 2016