Measurement of Inspiratory Flow Through Three Different Dry Powder Inhalation Devices Using In-Check TM in Children with Asthma
Authors:
- K.N. DesagerKDK.N. Desager
- J. GeldhofJGJ. Geldhof
- R. ClaesRCR. Claes
- W. De BackerWBW. De Backer
- 1 March 2006
- journal article
- research article
- Published by SAGE Publications in Pediatric Asthma, Allergy & Immunology
- Vol. 19 (1) , 6-13
Abstract
Inspiratory flow, generated by the patient through a dry powder inhalation device, determines particle size and dose delivered to the lungs. Children may have difficulty in generating the appropriate inspiratory flow to use these devices effectively. The aim is to evaluate inspiratory flow of children with asthma inhaling through In-CheckTM (Clement Clark International, Edinburgh Way, UK), equipped with adapters simulating resistance of three devices and to relate it to spirometric values and age. Forced spirometric parameters and inspiratory flow values through In-CheckTM were evaluated in stable children with asthma, aged 6 to 16 years. The inhalation devices tested were: AerolizerTM (Novartis, East Hanover, NJ), TurbuhalerTM (Astra Zeneca, Lund, Sweden), and DiskusTM (GlaxoSmithKline, Research Triangle Park, NC). Inspiratory flow through all devices exceeded 60 L/min in all but 1 child. More than half of the children (16/30) were able to generate a flow 120 L/min or more through AerolizerTM, 8 of 30 through DiskusTM, and 1 of 30 through TurbuhalerTM. For the TurbuhalerTM, inspiratory flows were significantly positively correlated with age. Inspiratory flows were significantly correlated with peak inspiratory flow (PIF) for all the devices. More children were able to obtain higher inspiratory flows through AerolizerTM. In-CheckTM is a valuable instrument to evaluate the inspiratory flow a child can generate through a dry powder inhalation device. (Pediatr Asthma Allergy Immunol 2006; 19[1]:6–13.)Keywords
This publication has 12 references indexed in Scilit:
- Variation of peak inspiratory flow through dry powder inhalers in children with stable and unstable asthmaPediatric Pulmonology, 2003
- Drug Delivery from the Turbuhaler and Nebuhaler Pressurized Metered Dose Inhaler to Various Age Groups of Children with AsthmaJournal of Aerosol Medicine, 1999
- Gender differences in respiratory, nasal and skin symptoms: 6-7 versus 13-14-year-old childrenActa Paediatrica, 1999
- Fine particle mass from the Diskus inhaler and Turbuhaler inhaler in children with asthmaEuropean Respiratory Journal, 1998
- Clinical effect of Diskus dry-powder inhaler at low and high inspiratory flow-rates in asthmatic childrenEuropean Respiratory Journal, 1998
- A comparison of the performance of two modern multidose dry powder asthma inhalersRespiratory Medicine, 1998
- Effect of Inhalation Flow Rate on the Dosing Characteristics of Dry Powder Inhaler (DPI) and Metered Dose Inhaler (MDI) ProductsJournal of Aerosol Medicine, 1996
- Drug delivery characteristics of Bricanyl TurbohalerTM dry powder inhalersInternational Journal of Pharmaceutics, 1995
- The Relationship Between Powder Inhaler Resistance and Peak Inspiratory Conditions in Healthy Volunteers — Implications for In Vitro TestingJournal of Aerosol Medicine, 1993
- Standardization of Spirometry—1987 UpdateAmerican Review of Respiratory Disease, 1987