The international standard ISO 15189 requires that medical laboratories estimate the uncertainty of their quantitative test results obtained from patients’ specimens. The standard does not provide details how and within which limits the measurement uncertainty should be determined. The most common concept for establishing permissible uncertainty limits is to relate them on biological variation defining the rate of false positive results or to base the limits on the state-of-the-art. The state-of-the-art is usually derived from data provided by a group of selected medical laboratories. The approach on biological variation should be preferred because of its transparency and scientific base. Hitherto, all recommendations were based on a linear relationship between biological and analytical variation leading to limits which are sometimes too stringent or too permissive for routine testing in laboratory medicine. In contrast, the present proposal is based on a non-linear relationship between biological and analytical variation leading to more realistic limits. The proposed algorithms can be applied to all measurands and consider any quantity to be assured. The suggested approach tries to provide the above mentioned details and is a compromise between the biological variation concept, the GUM uncertainty model and the technical state-of-the-art.Keywords
This publication has 9 references indexed in Scilit:
- Correction of an unexpected increasing trend in glucose measurements during 7 years recruitment to a cohort studyClinical Biochemistry, 2011
- Gross Overestimation of Total Allowable Error Based on Biological VariationClinical Chemistry, 2011
- Observed, unknown distributions of clinical chemical quantities should be considered to be log-normal: a proposalClinical Chemistry and Laboratory Medicine (CCLM), 2010
- A new statistical method for evaluating long-term analytical performance of laboratories applied to an external quality assessment scheme for flow cytometryClinical Chemistry and Laboratory Medicine (CCLM), 2010
- Cautions in the Adoption of Common Reference IntervalsClinical Chemistry, 2008
- A practical approach to instrument selection, evaluation, basic financial management and implementation in pathology and researchClinical Chemistry and Laboratory Medicine (CCLM), 2008
- A plea for intra-laboratory reference limits. Part 2. A bimodal retrospective concept for determining reference limits from intra-laboratory databases demonstrated by catalytic activity concentrations of enzymesClinical Chemistry and Laboratory Medicine (CCLM), 2007
- Setting Performance Goals and Evaluating Total Analytical Error for Diagnostic AssaysClinical Chemistry, 2002
- Biological and Analytic Components of Variation in Long-Term Studies of Serum Constituents in Normal SubjectsClinical Chemistry, 1970