Effects of High-Resolution CT of the Lung Using Partial Versus Full Reconstruction on Motion Artifacts and Image Noise
Authors:
- 1 September 2006
- journal article
- clinical trial
- Published by American Roentgen Ray Society in American Journal of Roentgenology
- Vol. 187 (3) , 618-622
Abstract
OBJECTIVE. The purpose of our study was to evaluate the effects of 0.3-second high-resolution CT (HRCT) of the lung using partial reconstruction on cardiac motion artifacts and image noise. SUBJECTS AND METHODS. Thirty-seven pairs of 0.3-second (partial reconstruction) and 0.75-second (full reconstruction) HRCT images were obtained for the lower lung zone during full-inspiration breath-holding. Imaging parameters other than temporal resolution were identical for each patient. Two radiologists visually graded motion artifacts of the cardiac border, bronchi, pulmonary vessels, and fissure in the left lung on a 4-point scale (with 4 indicating no artifacts). The maximum width of motion along the left cardiac border and the area percentage of motion artifacts in the left lung were calculated. Image noise in the air and lung was also determined. Cardiac motion artifacts and image noises were compared between the two sets of CT images. RESULTS. Visual grades for the cardiac border (4 ± 0), bronchi (3.8 ± 0.7), pulmonary vessels (3.6 ± 0.8), and fissure (3.9 ± 0.5) were higher for 0.3-second images than for 0.75-second images (1.7 ± 0.7, 2.0 ± 1.0, 1.6 ± 0.7, and 2.4 ± 0.9, respectively) (p < 0.001). The maximum width of motion along the left cardiac border (0.1 ± 0.5 mm) and the area percentage of motion artifacts in the left lung (6.7% ± 18.4%) were smaller for 0.3-second images than for 0.75-second images (4.5 ± 1.7 mm and 36.2% ± 20.9%, respectively) (p < 0.001). Image noises in the air (38.0 ± 9.2) and the lung (86.0 ± 23.1) were greater for 0.3-second images than for 0.75-second images (35.6 ± 9.6 and 76.0 ± 20.3, respectively) (p < 0.01). CONCLUSION. Compared with 0.75-second HRCT using full reconstruction, 0.3-second HRCT using partial reconstruction substantially reduces cardiac motion artifacts in the lung at the expense of increasing image noise.Keywords
This publication has 19 references indexed in Scilit:
- Electrocardiographically Gated 16-Section CT of the Thorax: Cardiac Motion SuppressionRadiology, 2004
- Thin-Section CT of the Lung: Does Electrocardiographic Triggering Influence Diagnosis?Radiology, 2003
- Lung at Thin-Section CT: Influence of Multiple-Segment Reconstruction on Image QualityRadiology, 2003
- Thoracic spiral CT: influence of subsecond gantry rotation on image quality.Radiology, 1998
- Motion artifacts on CT simulate bronchiectasisAmerican Journal of Roentgenology, 1988
- The double-fissure sign: a motion artifact on thin-section CT scans.Radiology, 1987
- Computed tomography of the pulmonary parenchyma. Part 2Journal of Thoracic Imaging, 1985
- Computed tomography of the pulmonary parenchyma. Part 1Journal of Thoracic Imaging, 1985
- Computed Tomography of BronchiectasisJournal of Computer Assisted Tomography, 1982
- The "twinkling star" sign: an aid in differentiating pulmonary vessels from pulmonary nodules on computed tomograms.Radiology, 1980