Abstract
A real-time PCR was designed for detection ofChlamydophila pneumoniaeandMycoplasma pneumoniaesuch that each pathogen could be detected in a single tube and differentiated using molecular beacons marked with different fluorochromes. This duplex PCR, targeting the P1 adhesion gene forM. pneumoniaeand theompAgene forC. pneumoniae, was compared with two conventional PCR assays targeting the 16S rRNA gene and theompAgene. A total of 120 clinical throat and nasopharyngeal swab samples were tested. DNA extraction was performed using an alkali denaturation/neutralization method, and real-time amplification, detection, and data analysis were performed using a Rotor-Gene 2000 real-time rotary analyzer (Corbett Life Science, Sydney, Australia). Using conventional PCR as a reference in an analysis of 120 samples, 13 of 14 samples positive forC. pneumoniaewere detected by the novel real-time PCR. In an analysis ofM. pneumoniae, 22 samples were positive in the conventional PCR and the novel assay detected 24 positive samples. When using the conventional PCR as a reference, sensitivity and specificity were 93% and 100%, respectively, forC. pneumoniaeand 100% and 98%, respectively, forM. pneumoniae. With an overall agreement of 98.8%, this suggests that performance of the new duplex real-time PCR is comparable to that of conventional PCR.

This publication has 39 references indexed in Scilit: