Pulmonary Vein Isolation With Single Pulse Irreversible Electroporation
- 1 October 2020
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation: Arrhythmia and Electrophysiology
- Vol. 13 (10) , 1083-1091
Abstract
Background - Irreversible electroporation (IRE) is a promising new non-thermal ablation technology for pulmonary vein (PV) isolation in patients with atrial fibrillation (AF). Experimental data suggest that IRE ablation produces large enough lesions without the risk of PV stenosis, artery, nerve or esophageal damage. This study aimed to investigate the feasibility and safety of single pulse IRE PV isolation in patients with AF. Methods - Ten patients with symptomatic paroxysmal or persistent AF underwent single pulse IRE PV isolation under general anesthesia. Three-dimensional reconstruction and electroanatomical voltage mapping (EnSite PrecisionTM, Abbott) of left atrium and PVs were performed using a conventional circular mapping catheter. PV isolation was performed by delivering non-arcing, non-barotraumatic 6 ms, 200 J direct current IRE applications via a custom non-deflectable 14-polar circular IRE ablation catheter with a variable hoop diameter (16-27 mm). A deflectable sheath (AgilisTM, Abbott) was used to maneuver the ablation catheter. A minimum of two IRE applications with slightly different catheter positions were delivered per vein to achieve circular tissue contact, even if PV potentials were abolished after the first application. Bidirectional PV isolation was confirmed with the circular mapping catheter and a post ablation voltage map. After a 30-minute waiting period, adenosine testing (30 mg) was used to reveal dormant PV conduction. Results - All 40 PVs could be successfully isolated with a mean of 2.4±0.4 IRE applications per PV. Mean delivered peak voltage and peak current were 2154 ± 59 V and 33.9 ± 1.6 A, respectively. No PV reconnections occurred during the waiting period and adenosine testing. No periprocedural complications were observed. Conclusions - In the 10 patients of this first-in-human study, acute bidirectional electrical PV isolation could be achieved safely by single pulse IRE ablation.Keywords
This publication has 34 references indexed in Scilit:
- ST-segment changes after direct current external cardioversion for atrial fibrillation. Incidence, characteristics and predictive factorsInternational Journal of Cardiology, 2011
- Feasibility of Electroporation for the Creation of Pulmonary Vein Ostial LesionsJournal of Cardiovascular Electrophysiology, 2011
- Transcranial Measurement of Cerebral Microembolic Signals During Endocardial Pulmonary Vein Isolation: Comparison of Three Different Ablation TechniquesJournal of Cardiovascular Electrophysiology, 2009
- Incidence and Time Course of Early Recovery of Pulmonary Vein Conduction after Catheter Ablation of Atrial FibrillationJournal of Cardiovascular Electrophysiology, 2007
- Phrenic Nerve Injury After Catheter Ablation: Should We Worry About This Complication?Journal of Cardiovascular Electrophysiology, 2006
- Comparison of Electrode Cooling Between Internal and Open Irrigation in Radiofrequency Ablation Lesion Depth and Incidence of Thrombus and Steam PopCirculation, 2006
- Catheter Ablation Without Fulguration: Design and Performance of a New SystemPacing and Clinical Electrophysiology, 1989
- Catheter ablation of the atrioventricular junction: a report of the percutaneous mapping and ablation registry.Circulation, 1984
- Catheter Technique for Closed-Chest Ablation of the Atrioventricular Conduction SystemNew England Journal of Medicine, 1982
- ST-segment elevation with elective DC cardioversion.Circulation, 1981