Early-stage Development of the CoRISMA Mechanical Circulatory Support (CMCS) System for Heart Failure Therapy
Authors:
- 22 July 2024
- journal article
- research article
- Published by Springer Nature in Cardiovascular Engineering and Technology
- Vol. 15 (6) , 667-678
Abstract
Purpose CoRISMA MCS Systems Inc (Hamden CT) is developing an innovative mechanical circulatory support system (CMCS) as a durable therapeutic option for heart failure (HF) patients. The CMCS system is comprised of an axial flow pump, non-contacting hydrodynamic bearings, and integrated DC motor designed to be fully implantable in a left atrial (LA) to aortic (Ao) configuration; this unloading strategy may be particularly beneficial for HF patients with preserved ejection fraction (HFpEF). The small (5.5 cm3), lightweight (20 g), and low power (5–7 W) device design should allow for a less invasive off-pump implant. We present early-stage engineering development and testing of the prototype CoRISMA pumps. Methods Computational fluid dynamics (CFD) modeling was performed to evaluate flow and shear in two impeller (3 blades, 0.5 mm thickness, 8.9 mm diameter, 0.15 mm gap, polished titanium) and diffusor (5 blades, polished titanium) candidate designs. Test apparatuses were custom built to expedite development of the impeller/diffuser designs and iteratively refine the CFD models. Two candidate impeller/diffusor designs were fabricated and tested in each of the two test apparatuses (n = 4 impeller/diffuser + test fixture configurations) in static mock flow loops (hydrodynamic H-Q curves, 3.5 cP glycerol solution at 37 °C), and in dynamic mock flow loops (hemodynamics, 3.5 cP glycerol solution at 37 °C) tuned to HF conditions (mean aortic pressure 50 mmHg, central venous pressure 15 mmHg, aortic flow 3.0 L/min, and heart rate 80 bpm). Results CFD predicted flows of 4.56 L/min and 4.82 L/min at 100 mmHg for impellers/diffusers 1 and 2, respectively. Impeller 2 required less torque to generate a 6% increase in fluidic flow, and the diffuser had a larger area of high pressure, indicative of lower friction, which likely contributed to the increased efficiency. Experimental testing for all four configurations in the static and dynamic mock loops met performance metrics as evidenced by generating 4.0–4.5 L/min flow against 70–76 mmHg pressure at 25,000 rpm and restoring hemodynamics in the dynamic mock flow loop (MAP = 80 mmHg, CVP = 0 mmHg, total flow = 5.5 L/min) from baseline simulated HF test conditions. Conclusion These results demonstrate proof-of-concept of the early engineering design and performance of the prototype CoRISMA pumps. Engineering specifications, challenges observed, and proposed solutions for the next design iteration were identified for the continued development of an effective, reliable, and safe LA-to-Ao CMCS system for HF patients. Current design plans are underway for incorporating a wireless energy transfer system for communication and power, eliminating the need for and complications associated with an external driveline, to achieve a fully-implantable system.Keywords
This publication has 25 references indexed in Scilit:
- Physiology and Clinical Utility of HeartMate Pump ParametersJournal of Cardiac Failure, 2021
- Can mechanical circulatory support be an effective treatment for HFpEF patients?Heart Failure Reviews, 2021
- Theoretical considerations for a left atrial pump in heart failure with preserved ejection fractionHeart Failure Reviews, 2021
- Left Ventricular Assist Device With a Left Atrial Inflow Cannula for Hypertrophic CardiomyopathyJACC: Case Reports, 2020
- The Physiology of Continuous-Flow Left Ventricular Assist DevicesJournal of Cardiac Failure, 2017
- Left Atrial Decompression Pump for Severe Heart Failure With Preserved Ejection FractionJACC: Heart Failure, 2015
- Energy Transmission and Power Sources for Mechanical Circulatory Support Devices to Achieve Total ImplantabilityThe Annals of Thoracic Surgery, 2014
- Clinical benefits of partial circulatory support in New York Heart Association Class IIIB and Early Class IV patientsEuropean Journal of Cardio-Thoracic Surgery, 2011
- Hemodynamic effects of partial ventricular support in chronic heart failure: Results of simulation validated with in vivo dataThe Journal of Thoracic and Cardiovascular Surgery, 2007
- Partial Support with a Centrifugal Left Ventricular Assist Device Reduces Myocardial Oxygen Consumption in Chronic, Ischemic Heart FailureJournal of Cardiac Failure, 2005