Long‐Term Evaluation of the Ventricular Defibrillation Energy Requirement
Authors:
- 1 September 1998
- journal article
- research article
- Published by Wiley in Journal of Cardiovascular Electrophysiology
- Vol. 9 (9) , 916-920
Abstract
Defibrillation energy requirements in patients with nonthoracotomy defibrillators may increase within several months after implantation. However, the stability of the defibrillation energy requirement beyond 1 year has not been reported. The purpose of this study was to characterize the defibrillation energy requirement during 2 years of clinical follow-up. Thirty-one consecutive patients with a biphasic nonthoracotomy defibrillation system underwent defibrillation energy requirement testing using a step-down technique (20, 15, 12, 10, 8, 6, 5, 4, 3, 2, and 1 J) during defibrillator implantation, and then 24 hours, 2 months, 1 year, and 2 years after implantation. The mean defibrillation energy requirement during these evaluations was 10.9+/-5.5 J, 12.3+/-7.3 J, 11.7+/-5.6 J, 10.2+/-4.0 J, and 11.7+/-7.4 J, respectively (P = 0.4). The defibrillation energy requirement was noted to have increased by 10 J or more after 2 years of follow-up in five patients. In one of these patients, the defibrillation energy requirement was no longer associated with an adequate safety margin, necessitating revision of the defibrillation system. There were no identifiable clinical characteristics that distinguished patients who did and did not develop a 10-J or more increase in the defibrillation energy requirement. The mean defibrillation energy requirement does not change significantly after 2 years of biphasic nonthoracotomy defibrillator system implantation. However, approximately 15% of patients develop a 10-J or greater elevation in the defibrillation energy requirement, and 3% may require a defibrillation system revision. Therefore, a yearly evaluation of the defibrillation energy requirement may be appropriate.Keywords
This publication has 21 references indexed in Scilit:
- Rise in Chronic Defibrillation Energy Requirements Necessitating Implantable Defibrillator Lead System RevisionPacing and Clinical Electrophysiology, 1997
- Chronic Rise in Monophasic Defibrillation Thresholds With a Transvenous Lead SystemThe American Journal of Cardiology, 1997
- Incidence of Lead System Malfunction Detected During Implantable Defibrillator Generator ReplacementPacing and Clinical Electrophysiology, 1996
- Increase in defibrillation threshold in non-thoracotorny implantable defibrillators using a biphasic waveformThe American Journal of Cardiology, 1995
- Early Postoperative Increase in Defibrillation Threshold with Nonthoracotomy System in HumansPacing and Clinical Electrophysiology, 1994
- Rise in chronic defibrillation thresholds in nonthoracotomy implantable defibrillator.Circulation, 1994
- Defibrillation Energy Requirements with Single Endocardial (Endotak™) LeadPacing and Clinical Electrophysiology, 1993
- Long‐Term Stability of Defibrillation Thresholds with Intrapericardial Defibrillator PatchesPacing and Clinical Electrophysiology, 1993
- Effects of chronic amiodarone therapy on defibrillation thresholdThe American Journal of Cardiology, 1992
- Success of chronic defibrillation and the role of antiarrhythmic drugs with the automatic implantable cardioverter/defibrillatorThe American Journal of Cardiology, 1987