LIQUID-CHROMATOGRAPHIC DETERMINATION OF 5-FLUOROCYTOSINE
Authors:
- J O MinersJMJ O Miners
- T FoenanderTFT Foenander
- D J BirkettDBD J Birkett
- 1 January 1980
- journal article
- research article
- Vol. 26 (1) , 117-119
Abstract
A high-performance liquid-chromatographic method for measuring 5-fluorocytosine in [human] plasma and CSF is reported. After deproteinization with trichloroacetic acid, the supernates are chromatographed on a reversed-phase (C18) column. Response to concentration is linear in the range of 5-200 mg/l, with UV detection at 276 nm. The assay requires only 0.1 ml of plasma, is reproducible, and may be performed in less than 12 min. 5-Fluorocytosine concentrations determined by this procedure correlated well with those obtained by spectrofluorometry, although the present method is more specific with no observable interference from coadministered amphotericin B and most other commonly encountered drugs, including salicylate. This method is applicable to the routine therapeutic monitoring of pediatric and adult patients and to pharmacokinetic studies.Keywords
This publication has 3 references indexed in Scilit:
- Comparison of liquid- and gas-liquid chromatographic assays of 5-fluorouracil in plasma.Clinical Chemistry, 1979
- Gas-chromatographic determination of 5-fluorocytosine in human serum.Clinical Chemistry, 1976
- Amphotericin B ToxicityAnnals of Internal Medicine, 1964