Foreign Body Reaction after Implantation of a Device for Intervertebral Assisted Motion
Open Access
- 1 January 2016
- journal article
- Published by Korean Neurosurgical Society in Journal of Korean Neurosurgical Society
- Vol. 59 (6) , 647-649
Abstract
The device for intervertebral assisted motion (DIAM) is a dynamic implant that consists of a silicone bumper enveloped by a polyethylene terephthalate (PET) fiber sack. Silicone and PET were used because of their biological inertness, but repetitive motion of the spine can cause wear on the implant nonetheless. The purpose of this study is to report a case of foreign body reaction (FBR) against a DIAM. A 72-year-old female patient presented with lower back pain and both legs radiating pain. She had undergone DIAM implantation at L4-5 for spinal stenosis 5 years previously. The intervertebral disc space of L4-5, where the DIAM was inserted, had collapsed and degenerative scoliosis had developed due to left-side collapse. MRI showed L3-4 thecal sac compression and left L4-5 foraminal stenosis. The patient underwent removal of the DIAM and instrumented fusion from L3 to L5. During surgery, fluid and granulation tissue were evident around the DIAM. Histopathology showed scattered wear debris from the DIAM causing chronic inflammation due to the resulting FBR. A FBR due to wear debris of a DIAM can induce a hypersensitivity reaction and bone resorption around the implant, causing it to loosen.Keywords
This publication has 10 references indexed in Scilit:
- Epidural application of spinal instrumentation particulate wear debris: a comprehensive evaluation of neurotoxicity using an in vivo animal modelJournal of Neurosurgery: Spine, 2013
- The Role of Silicone Granulomas in Recurrent Capsular ContracturePlastic and Reconstructive Surgery, 2013
- Posterior dynamic stabilization in the treatment of degenerative lumbar stenosis: validity of its rationaleJournal of Neurosurgery: Spine, 2013
- Paraspinal Textiloma After Posterior Lumbar Surgery: A Wolf in Sheep's ClothingWorld Neurosurgery, 2012
- Pseudotumors Associated with Total Hip ArthroplastyJournal of Bone and Joint Surgery, 2012
- Quantitative in vivo cytokine analysis at synthetic biomaterial implant sitesJournal of Biomedical Materials Research Part A, 2008
- Foreign body reaction due to polyethylene’s wear after implantation of an interspinal segmentArchives of orthopaedic and trauma surgery, 2007
- Interspinous Process SpacersJournal of the American Academy of Orthopaedic Surgeons, 2007
- Multinucleated giant cellsCurrent Opinion in Hematology, 2000
- Inflammatory Response to ImplantsASAIO Journal, 1988