Abstract
OBJECTIVE. The purpose of this study is to evaluate observer detection and volume measurement of small irregular solid artificial pulmonary nodules on 64-MDCT in an anthropomorphic thoracic phantom. MATERIALS AND METHODS. Forty in-house-made solid pulmonary nodules (lobulated and spiculated; actual volume, 5.1–88.4 mm3; actual CT densities, −51 to 157 HU) were randomly placed inside an anthropomorphic thoracic phantom with pulmonary vasculature. The phantom was examined on two 64-MDCT scanners, using a scan protocol as applied in lung cancer screening. Two independent blinded observers screened for pulmonary nodules. Nodule volume was evaluated semiautomatically using dedicated software and was compared with the actual volume using an independent-samples t test. The interscanner and interobserver agreement of volumetry was assessed using Bland-Altman analysis. RESULTS. Observer detection sensitivity increased along with increasing size of irregular nodules. Sensitivity was 100% when the actual volume was ...