Correction of the effects of source, source strength, and soft-tissue thickness on spine dual-photon absorptiometry measurements
Authors:
- 1 July 1989
- journal article
- research article
- Published by Springer Nature in Calcified Tissue International
- Vol. 44 (4) , 251-257
Abstract
Summary Dual photon absorptiometry measurements of the spine are subject to drift associated with source, source strength, and truncal thickness. This study was conducted to determine the extent to which this drift in bone mineral density (BMD) measurements can be improved by analysis of scans with a new software version, 08C, and by applying external standard or phantom corrections to scans analyzed with the older version, 08B. A phantom, consisting of human lumbar vertebrae embedded in acrylic, and five clear acrylic plates to simulate a soft-tissue thickness range of 15.2–27.9 cm, was measured on a Lunar Radiation Corp DP3 scanner over the life of a 153-gadolinium (Gd) source and scans analyzed with software versions 08B and 08C. Phantom BMD was lower with 08C at both high [0.012±0.002 (SEM) g/cm2,P2,PP=0.014). When the same scans were analyzed with 08C, the thickness-related effect was corrected whereas a small (0.011 g/cm2/year) source-strength effect persisted. The effects of source strength and truncal thickness on BMD were also evaluated in 40 humans scanned at two detector collimations to vary count rate. With 08B, mean BMD was 1% greater when measured with 8 than with 13 mm collimation (mean difference 0.011±0.003 g/cm2,P=0.001), whereas the version 08C, mean BMD was the same at the two collimations. Similarly, when phantom corrections were applied to the scans analyzed with 08B, the source strength effect was no longer significant. A truncal thickness effect, apparent in the 40 human scans analyzed with 08B, was not present with 08C. Finally, the phantom was scanned with three different Gd sources. With both 08B and 08C, BMD values were similar with two and significantly lower (by 0.012±0.002 g/cm2,P=0.011) with the third Gd source. Thus, with the new analysis software 08C, multiple thickness calibration is no longer needed, however, calibration with an external standard is still necessary.Keywords
This publication has 9 references indexed in Scilit:
- Precision of dual-photon absorptiometryCalcified Tissue International, 1988
- Forearm bone loss in hemiplegia: A model for the study of immobilization osteoporosisJournal of Bone and Mineral Research, 1988
- LONG-TERM PRECISION OF DUAL PHOTON-ABSORPTIOMETRY IN THE LUMBAR SPINE IN CLINICAL SETTINGS1988
- Precision error in dual-photon absorptiometry related to source age.Radiology, 1988
- Dual photon absorptimetric measurements of bone mineral density increase with source lifeCalcified Tissue International, 1987
- Dietary calcium intake and bone loss from the spine in healthy postmenopausal womenThe American Journal of Clinical Nutrition, 1987
- PRECISION OF DUAL PHOTON-ABSORPTIOMETRY MEASUREMENTS1986
- Changes in Bone Mineral Density of the Proximal Femur and Spine with AgingJournal of Clinical Investigation, 1982
- Total body bone mineral and lean body mass by dual-photon absorptiometryCalcified Tissue International, 1981