How physical infusion system parameters cause clinically relevant dose deviations after setpoint changes
Authors:
- Annemoon M. TimmermanATAnnemoon M. Timmerman
- Roland A. SnijderRSRoland A. Snijder
- Peter LucasPLPeter Lucas
- Martine C. LagerweijMLMartine C. Lagerweij
- Joris H. RadermacherJRJoris H. Radermacher
- Maurits K. KoningsMKMaurits K. Konings
- 1 January 2015
- journal article
- research article
- Published by Walter de Gruyter GmbH in Biomedical Engineering / Biomedizinische Technik
- Vol. 60 (4) , 365-376
Abstract
Multi-infusion therapy, in which multiple pumps are connected to one access point, is frequently used in patient treatments. This practice is known to cause dosing errors following setpoint changes in the drug concentrations that actually enter the patients. Within the Metrology for Drug Delivery Project, we analyzed and quantified the two main physical phenomena leading to these errors: the “push-out” effect and the system mechanical compliance. We compared the dosing errors of a three-pump system with two infusion sets, both with and without anti-reflux valves, using in vitro spectrophotometric experiments. Additionally, computer simulations were used to study the compliance effect separately. We found a start-up time of more than 1 h, and a dosing error following a setpoint increase of another pump for the low flow rate pump, corresponding to 0.5 μg noradrenaline delivered in 8 min. We showed that the dead volume inside the tubes and syringe compliance produce opposite deviations from the setpoint values in the actual drug output concentrations, making the net result hard to predict and often counterintuitive. We conclude that metrology on compliance and push-out effects could be used by infusion device manufacturers to successfully improve drug delivery performance and relevant standards for high-risk multi-infusion applications.Keywords
This publication has 21 references indexed in Scilit:
- A literature review on flow‐rate variability in neonatal IV therapyPediatric Anesthesia, 2012
- Incident hémodynamique majeur pendant une perfusion continue de noradrénaline : attention à la ligne de perfusion. Un pic hypertensif postopératoire évitable ?Annales Françaises d'Anesthésie et de Réanimation, 2012
- The Link Between Intravenous Multiple Pump Flow Errors and Infusion System Mechanical ComplianceAnesthesia & Analgesia, 2010
- An In Vitro Analysis of Central Venous Drug Delivery by Continuous Infusion: The Effect of Manifold Design and Port SelectionAnesthesia & Analgesia, 2009
- Impact of Multiaccess Infusion Devices on In Vitro Drug Delivery During Multi-Infusion TherapyAnesthesia & Analgesia, 2009
- The Delivery of Drugs to Patients by Continuous Intravenous Infusion: Modeling Predicts Potential Dose Fluctuations Depending on Flow Rates and Infusion System Dead VolumeAnesthesia & Analgesia, 2006
- The Impact of Carrier Flow Rate and Infusion Set Dead-Volume on the Dynamics of Intravenous Drug DeliveryAnesthesia & Analgesia, 2005
- Accurate continuous drug delivery at low infusion rate with a novel microvolumetric infusion pump (MVIP): pump design, evaluation and comparison to the current standardAnaesthesia, 2004
- Management of the changeover of inotrope infusions in childrenIntensive and Critical Care Nursing, 2004
- Practical considerations in the administration of intravenous vasoactive drugs in the critical care setting: the double pumping or piggyback technique?part oneIntensive and Critical Care Nursing, 2004