First-Trimester Detection of Surface Abnormalities: A Comparison of 2- and 3-Dimensional Ultrasound and 3-Dimensional Virtual Reality Ultrasound
Authors:
- 1 August 2014
- journal article
- research article
- Published by Springer Nature in Reproductive Sciences
- Vol. 21 (8) , 993-999
Abstract
The aim was to determine the diagnostic performance of 3-dimensional virtual reality ultrasound (3D_VR_US) and conventional 2- and 3-dimensional ultrasound (2D/3D_US) for first-trimester detection of structural abnormalities. Forty-eight first trimester cases (gold standard available, 22 normal, 26 abnormal) were evaluated offline using both techniques by 5 experienced, blinded sonographers. In each case, we analyzed whether each organ category was correctly indicated as normal or abnormal and whether the specific diagnosis was correctly made. Sensitivity in terms of normal or abnormal was comparable for both techniques (P = .24). The general sensitivity for specific diagnoses was 62.6% using 3D_VR_US and 52.2% using 2D/3D_US (P = .075). The 3D_VR_US more often correctly diagnosed skeleton/limb malformations (36.7% vs 10%; P = .013). Mean evaluation time in 3D_VR_US was 4:24 minutes and in 2D/3D_US 2:53 minutes (P < .001). General diagnostic performance of 3D_VR_US and 2D/3D_US apparently is comparable. Malformations of skeleton and limbs are more often detected using 3D_VR_US. Evaluation time is longer in 3D_VR_US.Keywords
This publication has 26 references indexed in Scilit:
- Fetal echocardiography at 11–13 weeks by transabdominal high‐frequency ultrasoundUltrasound in Obstetrics & Gynecology, 2011
- A model for a new pyramid of prenatal care based on the 11 to 13 weeks' assessmentPrenatal Diagnosis, 2011
- Embryonic delay in growth and development related to confined placental trisomy 16 mosaicism, diagnosed by I-Space Virtual RealityFertility and Sterility, 2008
- Reliability of three‐dimensional sonographic measurements in early pregnancy using virtual realityUltrasound in Obstetrics & Gynecology, 2008
- Detailed screening for fetal anomalies and cardiac defects at the 11–13‐week scanUltrasound in Obstetrics & Gynecology, 2006
- Screening for major structural abnormalities at the 11- to 14-week ultrasound scanAmerican Journal of Obstetrics and Gynecology, 2006
- Two‐stage ultrasonography in screening for fetal anomalies at 13–14 and 18–22 weeks of gestationActa Obstetricia et Gynecologica Scandinavica, 2004
- Learning curve in ultrasonographic screening for selected fetal structural anomalies in early pregnancyObstetrics & Gynecology, 2003
- Detection of fetal structural abnormalities at the 11–14 week ultrasound scanPrenatal Diagnosis, 2002
- Comparison of First and Second Trimester Screening for Fetal AnomaliesAnnals of the New York Academy of Sciences, 1998