Medical ultrasonic imaging technique (USI) is widely applied in clinical settings. However, it has not yet been applied for imaging and diagnosis of internal tissues of bones. This study aims to investigate and establish a multi-modal ultrasonic in bone imaging correction (MUBIC) paradigm combining ultrasonic (USI) and photoacoustic (PAI) advantages, and then to develop a USI and PAI bone imaging instrument (UPBII) by training and integrating with an attention-based Y-shaped network (AY-Net) for multi-modal ultrasonic imaging correction. The performance of MUBIC is demonstrated through both simulation and experiments. The results showed that contrast-to-noise ratio (CNR) of the corrected ultrasonic images increased from 3.2 decibels to 7.8 decibels, and the peak signal-to-noise ratio (PSNR) increased from 21.6 decibels to 34.3 decibels; while CNR of the corrected photoacoustic images increased from 2.8 decibels to 7.5 decibels, and PSNR increased from 22.6 decibels to 34.2 decibels. Moreover, the processing time of MUBIC was less than 1.0 second in both USI and PAI modalities. This study developed and applied UPBII to verify the performance of AY-Net, thereby demonstrating that MUBIC is a high-speed and effective paradigm for correcting USI and PAI of bones.