To evaluate short-term complications after femtosecond laser cuts for laser in situ keratomileusis (LASIK). Department of Ophthalmology, Goethe University, Frankfurt, Germany. Retrospective consecutive case series. Consecutive femtosecond laser cuts (1210 cuts in 624 patients) from LASIK performed with the Intralase FS60 between 2006 and 2014 were analyzed. The analysis looked at intraoperative and postoperative complications at 1 day, 1 week, and 1 month and factors with potential impact on the complication rate, such as ablation depth. No irregular, incomplete, or thin flaps, buttonholes, gas breakthrough, or free caps were observed. Intraoperatively, incomplete side cuts were observed in 4 cases (0.3%). Three of the 4 occurred during the first year of femtosecond laser usage and were explained by a software problem. Epithelial defects, limbal bleeding, and decentered flaps without negative impact on outcome were observed in 45 (3.7%), 12 (1.0%), and 5 (0.4%) eyes, respectively, on the treatment date. Diffuse lamellar keratitis (DLK) stage 1 was observed in 69 (5.7%), 36 (3.0%), and 3 (0.2%) eyes at 1 day, 1 week, and 1 month, respectively, and DLK stage 2 in 20 (1.7%), 6 (0.5%), and 2 (0.2%) eyes, respectively. An increase in total ablation depth resulted in a higher incidence of microstriae (P < .01 at 1 day, 1 week, and 1 month). An increase in the optical zone resulted in a significantly higher incidence of epithelial ingrowth at 1 day and 1 week (P < .01). No major cut-related complications were observed. The femtosecond laser is a safe option to create a flap in LASIK. Proprietary or commercial disclosures are listed after the references.Keywords
This publication has 28 references indexed in Scilit:
- Efficacy, safety, predictability, contrast sensitivity, and aberrations after femtosecond laser lenticule extractionJournal of Cataract & Refractive Surgery, 2014
- Safety, efficacy, and predictability of laser in situ keratomileusis to correct myopia or myopic astigmatism with a 750 Hz scanning-spot laser systemJournal of Cataract & Refractive Surgery, 2014
- Laser in situ keratomileusis flap complications using mechanical microkeratome versus femtosecond laser: Retrospective comparisonJournal of Cataract & Refractive Surgery, 2010
- Corneal architecture of femtosecond laser and microkeratome flaps imaged by anterior segment optical coherence tomographyJournal of Cataract & Refractive Surgery, 2009
- Flap thickness reproducibility in laser in situ keratomileusis with a femtosecond laser: Optical coherence tomography measurementJournal of Cataract & Refractive Surgery, 2008
- Femtosekundenlaser für die refraktive HornhautchirurgieDer Ophthalmologe, 2006
- Reproducibility of Flap Thickness With IntraLase FS and Moria LSK-1 and M2 MicrokeratomesJournal of Refractive Surgery, 2006
- Laser in situ keratomileusis: Literature review of a developing techniqueJournal of Cataract & Refractive Surgery, 1998
- A Corneal Flap Technique for Laser In Situ KeratomileusisArchives of Ophthalmology (1950), 1991
- Laser in situ keratomileusisLasers in Surgery and Medicine, 1990