Out With the Old and in With the New? Factors Involved in Migration of Older and Newer Generation Stereotactic Breast Biopsy Markers
- 13 March 2021
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Breast Imaging
- Vol. 3 (4) , 465-472
Abstract
The purpose of this study was to investigate four commonly used stereotactic biopsy markers, two older and two newer generation, assessing percentage migration and factors influencing migration distance. This was an IRB–approved retrospective review of upright stereotactic breast biopsies from May 2018 to May 2020 involving either older (Cork, Hourglass) or newer (Vision, X-shaped) generation markers. Markers were assessed for migration rate by two-sample Z-test and migration distance by analysis of variance. Univariate analysis was used to assess relationships between marker type and generation, patient characteristics, breast composition and thickness, procedure techniques, trainee involvement, and complications, correlating with migration distance. Multivariable analysis was performed for variables with P-value < 0.1 on univariate analysis. Tukey’s test was used to compare all groups (P < 0.05). A total of 732 stereotactic biopsies were performed with 508 using a Cork, Hourglass, Vision, or X-shaped marker. Overall migration rate was 181/508 (35.6%) with no difference between markers. Breast thickness and density were negatively associated with migration distance in univariate analysis. Older marker migration distance was greater than newer (2.6 cm vs 1.9 cm, respectively), which was significant after adjusting for breast thickness and density (P = 0.037). Density was a significant factor in migration distance, comparing fatty to nonfatty breasts (P < 0.05) in univariate analysis. No difference in migration rate was seen between the four biopsy markers. Vision and X-shaped markers demonstrate lower migration distance than Cork and Hourglass in multivariate analysis. There is an inverse relationship between breast density and marker migration distance.Keywords
This publication has 13 references indexed in Scilit:
- Effect of Mammography on Marker Clip Migration After Stereotactic-Guided Core Needle Breast BiopsyCurrent Problems in Diagnostic Radiology, 2017
- Stereotactic core needle breast biopsy marker migration: An analysis of factors contributing to immediate marker migrationEuropean Radiology, 2017
- Clip placement after an 11-gauge vacuum-assisted stereotactic breast biopsy: correlation between breast thickness and clip movementBreast Cancer, 2011
- Recognizing Pitfalls in Early and Late Migration of Clip Markers after Imaging-guided Directional Vacuum-assisted BiopsyRadioGraphics, 2004
- Metallic Marker Placement After Stereotactic Core Biopsy of Breast Calcifications:Comparison of Two Clips and Deployment TechniquesAmerican Journal of Roentgenology, 2003
- Accuracy of a Collagen-Plug Biopsy Site Marking Device Deployed After Stereotactic Core Needle Breast BiopsyAmerican Journal of Roentgenology, 2003
- Clip migration in stereotactic biopsyThe American Journal of Surgery, 2002
- Clip Migration after 11-gauge Vacuum-assisted Stereotactic Biopsy: Case ReportRadiology, 2002
- Metallic Clip Deployment during Stereotactic Breast Biopsy: Retrospective AnalysisRadiology, 2001
- Tissue marking clip for stereotactic breast biopsy: initial placement accuracy, long-term stability, and usefulness as a guide for wire localization.Radiology, 1997