Remote Glucose Monitoring in Camp Setting Reduces the Risk of Prolonged Nocturnal Hypoglycemia
Authors:
- 1 January 2014
- journal article
- research article
- Published by SAGE Publications in Diabetes Technology & Therapeutics
- Vol. 16 (1) , 1-7
Abstract
Objective: This study tested the feasibility and effectiveness of remote continuous glucose monitoring (CGM) in a diabetes camp setting. Subjects and Methods: Twenty campers (7–21 years old) with type 1 diabetes were enrolled at each of three camp sessions lasting 5–6 days. On alternating nights, 10 campers were randomized to usual wear of a Dexcom (San Diego, CA) G4™ PLATINUM CGM system, and 10 were randomized to remote monitoring with the Dexcom G4 PLATINUM communicating with the Diabetes Assistant, a cell phone platform, to allow wireless transmission of CGM values. Up to 15 individual graphs and sensor values could be displayed on a single remote monitor or portable tablet. An alarm was triggered for values P =0.078). Remote monitoring significantly decreased prolonged hypoglycemic events, eliminating all events <50 mg/dL lasting longer than 30 min as well as all events <70 mg/dL lasting more than 2 h. Conclusions: Remote monitoring is feasible at diabetes camps and effective in reducing the risk of prolonged nocturnal hypoglycemia. This technology will facilitate forthcoming studies to evaluate the efficacy of automated closed-loop systems in the camp setting.Keywords
This publication has 25 references indexed in Scilit:
- Continuous glucose monitoring system during physical exercise in adolescents with type 1 diabetesActa Paediatrica, 2011
- Glucose Requirements to Maintain Euglycemia after Moderate-Intensity Afternoon Exercise in Adolescents with Type 1 Diabetes Are Increased in a Biphasic MannerThe Journal of Clinical Endocrinology & Metabolism, 2007
- Defective Awakening Response to Nocturnal Hypoglycemia in Patients with Type 1 Diabetes MellitusPLOS Medicine, 2007
- Efficacy of Continuous Real-Time Blood Glucose Monitoring During and After Prolonged High-Intensity Cycling Exercise: Spinning with a Continuous Glucose Monitoring SystemDiabetes Technology & Therapeutics, 2006
- Impact of Exercise on Overnight Glycemic Control in Children with Type 1 Diabetes MellitusThe Journal of Pediatrics, 2005
- Response to Nocturnal Alarms Using a Real-Time Glucose SensorDiabetes Technology & Therapeutics, 2005
- Mini-dose glucagon is effective at diabetes campThe Journal of Pediatrics, 2004
- Use of the Cygnus GlucoWatch Biographer at a Diabetes CampPediatrics, 2004
- Dead-in-bed syndrome in diabetes mellitusThe Lancet, 1997
- Unexplained Deaths of Type 1 Diabetic PatientsDiabetic Medicine, 1991