Remote Glucose Monitoring in Camp Setting Reduces the Risk of Prolonged Nocturnal Hypoglycemia

Authors:

Abstract
Objective: This study tested the feasibility and effectiveness of remote continuous glucose monitoring (CGM) in a diabetes camp setting. Subjects and Methods: Twenty campers (7–21 years old) with type 1 diabetes were enrolled at each of three camp sessions lasting 5–6 days. On alternating nights, 10 campers were randomized to usual wear of a Dexcom (San Diego, CA) G4™ PLATINUM CGM system, and 10 were randomized to remote monitoring with the Dexcom G4 PLATINUM communicating with the Diabetes Assistant, a cell phone platform, to allow wireless transmission of CGM values. Up to 15 individual graphs and sensor values could be displayed on a single remote monitor or portable tablet. An alarm was triggered for values P =0.078). Remote monitoring significantly decreased prolonged hypoglycemic events, eliminating all events <50 mg/dL lasting longer than 30 min as well as all events <70 mg/dL lasting more than 2 h. Conclusions: Remote monitoring is feasible at diabetes camps and effective in reducing the risk of prolonged nocturnal hypoglycemia. This technology will facilitate forthcoming studies to evaluate the efficacy of automated closed-loop systems in the camp setting.