Stainless Steel-Vinylidene Fluoride Resin Complete Hip Prosthesis—Evaluation in the Dog

Authors:

Abstract
SUMMARY: A complete hip prosthesis, consisting of a stainless steel acetabular cup, vinylidene fluoride resin (vfr) cup liner, and stainless steel femoral head, neck, and intramedullay shaft, was installed in 13 dogs. Evaluation was based on function, range of hip motion, stability of the prosthetic device, radiographic changes, and gross and histopathologic changes in the periarticular tissues. Excellent to good results were obtained in 5 dogs; fair to poor results were noted in 5. Two dogs were withdrawn from the test after 11 days, and 1 dog developed osteomyelitis and was euthanatized 4 months postoperatively. Variable degrees of soft tissue and bony reaction developed in all dogs. Failure of the vfr liner precludes the use of vfr on a weight-bearing surface in a prosthetic device where it will be subjected to abrasion. This prosthesis generally failed to produce results superior to those observed in dogs with coxofemoral pseudoarthrosis.

This publication has 7 references indexed in Scilit: