Case series of late lead dislodgement of Medtronic SelectSecure 3830 pacing leads in growing paediatric patients
Open Access
- 12 January 2021
- journal article
- research article
- Published by Oxford University Press (OUP) in European Heart Journal - Case Reports
- Vol. 5 (2) , ytaa545
Abstract
Background : The SelectSecure lumenless 3830 pacing lead is often considered to be the pacing lead of choice for transvenous pacing in children because of its small diameter, lead strength, and reliable long-term sensing and pacing characteristics. One of the potential long-term pitfalls of a sturdy pacing lead is relative retraction with growth in children resulting in late lead dislodgement.Case summary : We report two cases of late SelectSecure 3830 lead dislodgement at 11.8 years (Case 1) and 8.8 years (Case 2), respectively, post the initial implantation. Case 1 was diagnosed with congenital complete heart block (CHB) at 9 months old when he presented with unconfirmed diphtheria infection. Case 2 was diagnosed with CHB at 14 weeks of age with positive maternal anti-Ro antibodies. Both patients underwent implantation of a transvenous permanent pacemaker implantation with Medtronic SelectSecure 3830 lead due to symptomatic bradycardia. Apart from a pulse generator change at 8.5 years (Case 1) and 7 years (Case 2), respectively, post-implant due to normal battery depletion, both patients are well in the interim.Discussion : As part of the pacemaker follow-up for rapidly growing children, we recommend more frequent surveillance of lead ‘tautness’ by chest radiography especially in children with CHB with no underlying heart rhythm.Keywords
This publication has 21 references indexed in Scilit:
- Selective site pacing in paediatric patients: how to recreate a new loop in the right atrium to follow body growthEP Europace, 2011
- Performance of the Lumenless 4.1‐Fr Diameter Pacing Lead Implanted at Alternative Pacing Sites in Congenital Heart: A Chronic 5‐Year ComparisonPacing and Clinical Electrophysiology, 2010
- Initial Experience of Pacing with a Lumenless Lead System in Patients with Congenital Heart DiseasePacing and Clinical Electrophysiology, 2009
- Noncatheter‐Based Delivery of a Single‐Chamber Lumenless Pacing Lead in Small ChildrenPacing and Clinical Electrophysiology, 2007
- Twenty years of paediatric cardiac pacing: 515 pacemakers and 480 leads implanted in 292 patientsEP Europace, 2006
- Age, Size, and Lead Factors Alone Do Not Predict Venous Obstruction in Children and Young Adults with Transvenous Lead SystemsJournal of Cardiovascular Electrophysiology, 2006
- Inferior Vena Cava Loop of the Implantable Cardioverter Defibrillator Endocardial Lead: A Possible Solution to the Growth Problem in Pediatric ImplantationPacing and Clinical Electrophysiology, 2000
- Cardiac strangulation, a rare complication of epicardial pacemaker leads during growth.Heart, 1997
- Permanent cardiac pacing in children: morbidity and efficacy of follow-upInternational Journal of Cardiology, 1991
- Transvenous pacemakers in children: Relation of lead length to anticipated growthThe Annals of Thoracic Surgery, 1991