New Dental Implant with 3D Shock Absorbers and Tooth-Like Mobility—Prototype Development, Finite Element Analysis (FEA), and Mechanical Testing
Authors:
- Avram ManeaAMAvram Manea
- Grigore BaciutGBGrigore Baciut
- Mihaela BaciutMBMihaela Baciut
- Dumitru PopDPDumitru Pop
- Dan Sorin ComsaDCDan Sorin Comsa
- Ovidiu BuigaOBOvidiu Buiga
- Veronica TrombitasVeronica Trombitas
- Horatiu ColosiHoratiu Colosi
- Ileana MitreIMIleana Mitre
- Roxana BordeaRoxana Bordea
- Marius ManoleMMMarius Manole
- Manuela LenghelMLManuela Lenghel
Open Access
Abstract
Background: Once inserted and osseointegrated, dental implants become ankylosed, which makes them immobile with respect to the alveolar bone. The present paper describes the development of a new and original implant design which replicates the 3D physiological mobility of natural teeth. The first phase of the test followed the resistance of the implant to mechanical stress as well as the behavior of the surrounding bone. Modifications to the design were made after the first set of results. In the second stage, mechanical tests in conjunction with finite element analysis were performed to test the improved implant design. Methods: In order to test the new concept, 6 titanium alloy (Ti6Al4V) implants were produced (milling). The implants were fitted into the dynamic testing device. The initial mobility was measured for each implant as well as their mobility after several test cycles. In the second stage, 10 implants with the modified design were produced. The testing protocol included mechanical testing and finite element analysis. Results: The initial testing protocol was applied almost entirely successfully. Premature fracturing of some implants and fitting blocks occurred and the testing protocol was readjusted. The issues in the initial test helped design the final testing protocol and the new implants with improved mechanical performance. Conclusion: The new prototype proved the efficiency of the concept. The initial tests pointed out the need for design improvement and the following tests validated the concept.Keywords
Funding Information
- ‘Iuliu Hațieganu’ University of Medicine and Pharmacy, Cluj-Napoca, Romania (PCD 7690/67/15.04.2016, 22714/5/06.10.2011)
This publication has 62 references indexed in Scilit:
- Corrosion Behavior of Ti6Al4V Coated with SiOx by PECVD TechnologyKey Engineering Materials, 2013
- Measurement of tooth and implant mobility under physiological loading conditionsAnnals of Anatomy - Anatomischer Anzeiger, 2012
- Evaluation of test protocol variables for dental implant fatigue researchDental Materials, 2009
- Effect of different mucosa thickness and resiliency on stress distribution of implant-retained overdentures-2D FEAComputer Methods and Programs in Biomedicine, 2008
- A 16–year Study of the Microgap Between 272 Human Titanium Implants and Their AbutmentsJournal of Oral Implantology, 2005
- Biomechanical aspects of fixed bridgework supported by natural teeth and endosseous implantsPeriodontology 2000, 1994
- Recent results concerning physiological tooth movement and anterior guidanceJournal of Oral Rehabilitation, 1981
- Measurement of the Physiological Mobility of Individual Teeth in an Axial DirectionJournal of Dental Research, 1960
- Tooth Mobility: The Measuring Method. Initial and Secondary Tooth MobilityJournal of Periodontology, 1954
- Periodontometry, a method for measuring tooth mobilityOral Surgery, Oral Medicine, Oral Pathology, 1951