Long-term stimulation by active epiretinal implants in normal and RCD1 dogs
Authors:
- James D WeilandJWJames D Weiland
- Gildo FujiiGFGildo Fujii
- Brian V MechBMBrian V Mech
- Manjunatha MahadevappaManjunatha Mahadevappa
- Robert GreenbergRGRobert Greenberg
- Roberto RoizenblattRRRoberto Roizenblatt
- Guanting QiuGQGuanting Qiu
- Laurie LaBreeLLLaurie LaBree
- Xiaopeng WangXWXiaopeng Wang
- David HintonDHDavid Hinton
- Mark S HumayunMHMark S Humayun
- 23 February 2005
- journal article
- research article
- Published by IOP Publishing in Journal of Neural Engineering
- Vol. 2 (1) , S65-S73
Abstract
An epiretinal prosthesis, consisting of an extraocular microelectronic stimulator and an intraocular electrode array, was implanted in one eye of three blind and three sighted dogs. Three dogs (2 blind, 1 normal) were stimulated for 120 days, and two dogs (both normal) for 60 and 103 days respectively for 8-10 h/day at levels of 0.1 mC cm(-2) and 0.05 mC cm(-2), with each stimulus level presented to half of the array. One blind dog was kept as an inactive implant control. During the study period, electroretinograms (ERG) and fundus photographs were recorded. At the end of the study period, the dogs were sacrificed and histological and morphometric evaluation was made of the retina. No inflammatory reaction, neovascularization or hemorrhage was observed during the follow-up examinations. ERGs were unchanged. Stimulus levels used were of sufficient amplitude to elicit cortical evoked potentials. Histological evaluation showed no inflammatory infiltrates or changes in retina morphometry related to electrical stimulation when compared to the unstimulated control eye. Morphometric analysis revealed no consistent differences relating to electrical stimulation. In summary, chronic electrical stimulation of the dog retina at up to 0.1 mC cm(-2) with an epiretinal prosthesis does not appear to adversely affect the retina.Keywords
This publication has 25 references indexed in Scilit:
- Arrays for chronic functional microstimulation of the lumbosacral spinal cordIEEE Transactions on Neural Systems and Rehabilitation Engineering, 2004
- Perceptual Efficacy of Electrical Stimulation of Human Retina with a Microelectrode Array during Short-Term Surgical TrialsInvestigative Ophthalmology & Visual Science, 2003
- Immunohistochemical Studies of the Retina Following Long-term Implantation with Subretinal Microphotodiode ArraysExperimental Eye Research, 2001
- SUCCESSFUL LONG-TERM IMPLANTATION OF ELECTRICALLY INACTIVE EPIRETINAL MICROELECTRODE ARRAYS IN RABBITSRetina, 1999
- The Responses of Mammalian Spinal Axons to an Applied DC Voltage GradientExperimental Neurology, 1997
- Preservation of the Inner Retina in Retinitis PigmentosaArchives of Ophthalmology (1950), 1997
- Molecular diagnostic tests for ascertainment of genotype at the rod cone dysplasia 1 (rcd1) locus in Irish settersCurrent Eye Research, 1995
- Consequences of chronic extracochlear electrical stimulation in neonatally deafened catsHearing Research, 1995
- Morphometric Analysis of Macular Photoreceptors and Ganglion Cells in Retinas With Retinitis PigmentosaArchives of Ophthalmology (1950), 1992
- Chronic intracochlear electrical stimulation in the neonatally deafened cat. I: Expansion of central representationHearing Research, 1990