A flexible high voltage biphasic current-controlled stimulator
Authors:
- 1 November 2006
- conference paper
- Published by Institute of Electrical and Electronics Engineers (IEEE)
- No. 21634025,p. 206-209
Abstract
The efficiency of implantable stimulation systems depends in large proportion of the electrode-tissue interface condition and in the choice of appropriate stimuli waveform. In this paper, we present an integrated flexible stimuli generator dedicated for high current stimulation in large electrode-tissue contact impedance. The system, powered by a single 3.3 V supply level, includes an integrated high voltage generator in order to supplies the output stage with -10 V/+10 V. High voltage parts are implemented in a 9 mm2 chip using a 0.8 um high voltage process. For area and power efficiency concern, the control unit of the whole device is implemented in a standard 0.18 um technology and use 1.96 mm2 of silicon area. Preliminary results indicate that the stimuli generator can deliver a stimulating current of 2 mA in electrode-tissue impedance as high as 9.125 kOmega.Keywords
This publication has 5 references indexed in Scilit:
- An implantable asic for neural stimulation2005
- A Compact Large Voltage-Compliance High Output-Impedance Programmable Current Source for Implantable MicrostimulatorsIEEE Transactions on Biomedical Engineering, 2004
- Programmable current source dedicated to implantable microstimulators2002
- Neural and connective tissue response to long-term implantation of multiple contact nerve cuff electrodesJournal of Biomedical Materials Research, 2000
- MOS charge pumps for low-voltage operationIEEE Journal of Solid-State Circuits, 1998