A Comparison of Proximal Tibia, Distal Femur, and Proximal Humerus Infusion Rates Using the EZ-IO Intraosseous Device on the Adult Swine (Sus scrofa) Model
Authors:
- Julio LairetJulio Lairet
- Vikhyat BebartaVBVikhyat Bebarta
- Kimberly LairetKLKimberly Lairet
- Robert KacprowiczRKRobert Kacprowicz
- Christopher LawlerCLChristopher Lawler
- Rebecca PitottiRPRebecca Pitotti
- Anneke BushABAnneke Bush
- James KingJKJames King
- 18 January 2013
- journal article
- research article
- Published by Taylor & Francis in Prehospital Emergency Care
- Vol. 17 (2) , 280-284
Abstract
Objective. We sought to compare the flow rates of the proximal tibia, the distal femur, and the proximal humerus using high pressure (greater than 300 mmHg) through an intraosseous (IO) infusion needle in an adult swine model. Methods. We performed a prospective interventional study in 11 swine. After placement of central vein and arterial lines, blood was removed via the central line until the animal's mean arterial pressure decreased 25% from the recorded baseline. We inserted a 25-mm IO needle into the proximal tibia and 45-mm needles into the distal femur and proximal humerus. All extremities were utilized in each study animal. We infused normal saline at each site for 10 minutes with a pressure bag inflated to the highest achievable pressure (greater than 300 mmHg) as measured at the infusion site with a calibrated portable inline pressure gauge. We measured the volume of normal saline remaining and we calculated infusion rates for each site. We then compared infusion flow rates between the three locations. Statistical analysis and comparison of the infusion rates of all three study arms were as performed using analysis of variance (ANOVA). Results. The average weight of the swine was 71 kg (range 64–84 kg). Successful placement of the IO needle was confirmed at all three sites. The mean infusion flow rate was 213 mL/min (standard deviation [SD] 53.2 mL/min) for the proximal humerus, 138 mL/min (SD 65.3 mL/min) for the distal femur, and 103 mL/min (SD 48.1 mL/min) for the proximal tibia (p < 0.001). The flow rate through the proximal humerus was statistically greater than that for the proximal tibia and the distal femur (p < 0.001). The flow rates through the proximal tibia and distal femur were similar. Conclusion. The humerus is a suitable alternative site for IO placement, with a potential for higher flow rates than the proximal tibia and distal femur when resuscitating a patient. Key words: intraosseous; prehospital; resuscitation; tactical combat casualty care; emergency medical servicesKeywords
This publication has 16 references indexed in Scilit:
- Comparison of First-Attempt Success Between Tibial and Humeral Intraosseous Insertions During Out-of-Hospital Cardiac ArrestPrehospital Emergency Care, 2011
- Proximal Humerus Intraosseous Infusion: A Preferred Emergency Venous AccessJournal of Trauma: Injury, Infection & Critical Care, 2009
- Guidelines for Prehospital Fluid Resuscitation in the Injured PatientJournal of Trauma: Injury, Infection & Critical Care, 2009
- Is the intraosseous access route fast and efficacious compared to conventional central venous catheterization in adult patients under resuscitation in the emergency department? A prospective observational pilot studyPatient Safety in Surgery, 2009
- COMPARISON OF INTRAOSSEOUS, CENTRAL, AND PERIPHERAL ROUTES OF CRYSTALLOID INFUSION FOR RESUSCITATION OF HEMORRHAGIC SHOCK IN A SWINE MODEL1993
- Comparison of fluid infusion rates among peripheral intravenous and humerus, femur, malleolus, and tibial intraosseous sites in normovolemic and hypovolemic pigletsAnnals of Emergency Medicine, 1993
- Comparison of intraosseous and intravenous delivery of hypertonic saline/dextran in anesthetized, euvolemic pigsAnnals of Emergency Medicine, 1992
- Comparison of intraosseous and intravenous routes of anticonvulsant administration in a porcine modelAnnals of Emergency Medicine, 1989
- Analysis of growth plate abnormalities following intraosseous infusion through the proximal tibial epiphysis in pigsAnnals of Emergency Medicine, 1988
- The effect of intraosseous sodium bicarbonate on bone in swineAnnals of Emergency Medicine, 1987