Microneedle Insertion Force Reduction Using Vibratory Actuation
- 1 September 2004
- journal article
- research article
- Published by Springer Nature in Biomedical Microdevices
- Vol. 6 (3) , 177-182
Abstract
The effect of vibratory actuation on microneedle insertion force was investigated. Hollow micro hypodermic injection needles were fabricated by a two-wafer polysilicon micromolding process. A vibratory actuator operating in the kHz range was coupled with the hypodermic microneedles. The force to insert microneedles into excized animal tissue was measured with a load cell. Results showed a greater than 70% reduction in microneedle insertion force by using vibratory actuation. The application of vibratory actuation provides a promising method to precisely control the microneedle insertion forces to overcome microneedle structural material limitations, minimize insertion pain, and enhance the efficiency of drug delivery.Keywords
This publication has 10 references indexed in Scilit:
- Design, Fabrication and Testing of a MEMS Syringe2002
- Lack of Pain Associated with Microfabricated MicroneedlesAnesthesia & Analgesia, 2001
- Microfabricated Polysilicon Microneedles for Minimally Invasive Biomedical DevicesBiomedical Microdevices, 2000
- Microfabricated Microneedles for Gene and Drug DeliveryAnnual Review of Biomedical Engineering, 2000
- Active Microneedles with Integrated Functionality2000
- Microelectromechanical Systems Technology to Deliver InsulinDiabetes Technology & Therapeutics, 1999
- Micromachined needle arrays for drug delivery or fluid extractionIEEE Engineering in Medicine and Biology Magazine, 1999
- Polymolding: Two Wafer Polysilicon Micromolding of Closed-Flow Passages for Microneedles and Microfluidic Devices1998
- A MULTICHANNEL NEURAL PROBE FOR SELECTRIVE CHEMICAL DELIVERY AT THE CELLULAR LEVEL1994
- Ultrasound induced delivery of peptidesJournal of Controlled Release, 1993