Abstract
Wearable monitors for measuring vital signs such as blood pressure will greatly impact the medical field. This work presents a millimeter-wave, radar-based system for performing accurate measurements of arterial pulse waveforms without contacting the region that is pulsing. Electromagnetic and radar-system simulation models are utilized to demonstrate the viability and safety of this approach. This is followed by hardware/software implementation and a study on 12 human subjects. Measured radial arterial waveforms exhibit signal strengths that are well above the noise floor of the system and a morphology that would be expected in an arterial pulse. Finally, comparison of the radar-based signals with a reference tonometer indicates a strong correlation between waveforms, as well as similar spectral signatures. The results observed suggest a millimeter-wave based approach for arterial pulse detection is very promising for future applications in pulse wave analysis and pulse transit time measurement for blood pressure tracking.