Extension of the reconstruction field of view and truncation correction using sinogram decomposition
- 18 April 2007
- journal article
- radiation imaging-physics
- Published by Wiley in Medical Physics
- Vol. 34 (5) , 1593-1604
Abstract
We propose a novel truncation correction algorithm that completes unmeasured data outside of the scan field of view, which allows extending the reconstruction field of view. When a patient extends outside the detector coverage the projection data are transversely truncated, which causes severe artifacts. The proposed method utilizes the idea of sinogram decomposition, where we consider sinogram curves corresponding to image points outside the field of view. We propose two ways to estimate the truncated data, one based on the minimum value along the sinogram curve, and the other based on the data values near the edge of truncation. Both estimation methods are combined to achieve uniform image quality improvement from the edge of truncation to the outer side of the extended region. In our evaluation with simulated and real projection data we compare the proposed method with existing methods and investigate the dependence on the amount of truncation. The evaluation shows that the proposed method handles cases when truncation is present on both sides of the detector, or when a high‐contrast object is located outside the field of view.Keywords
This publication has 20 references indexed in Scilit:
- Helical cone beam CT with an asymmetrical detectorMedical Physics, 2005
- Image reconstruction in regions-of-interest from truncated projections in a reduced fan-beam scanPhysics in Medicine & Biology, 2004
- Quantitative reconstruction from truncated projections in classical tomographyIEEE Transactions on Nuclear Science, 2004
- A two-step Hilbert transform method for 2D image reconstructionPhysics in Medicine & Biology, 2004
- PET/CT scanners: A hardware approach to image fusionSeminars in Nuclear Medicine, 2003
- Technical note: Sinogram merging to compensate for truncation of projection data in tomotherapy imagingMedical Physics, 2002
- Image reconstruction from fan-beam projections on less than a short scanPhysics in Medicine & Biology, 2002
- X‐ray micro‐CT with a displaced detector arrayMedical Physics, 2002
- Efficient correction for CT image artifacts caused by objects extending outside the scan field of viewMedical Physics, 2000
- Cone-beam CT for radiotherapy applicationsPhysics in Medicine & Biology, 1995