Abstract
This study compared the speech recognition performance of 12 hearing-impaired listeners fit with three commercially available behind-the-ear hearing aids in both directional and omnidirectional modes. One digitally programmable analog and two "true digital" hearing aids were selected as test instruments. Testing was completed in both "living room" and anechoic room environments. Speech recognition was examined using modified forms of the Hearing in Noise Test and the Nonsense Syllable Test. The single competing stimuli of these tests were replaced with five uncorrelated competing sources. Results revealed a significant speech recognition in noise advantage for all directional hearing aids in comparison to their omnidirectional counterparts. Maximum performance of the directional hearing aids did not significantly vary across circuit type, suggesting that processing differences did not affect maximum directional hearing aid performance. In addition, the results suggest that performance in one reverberant environment cannot be used to accurately predict performance in an environment with differing reverberation. Abbreviations: A = Phonak Piconet P2 AZ™ omnidirectional; AD = Phonak Piconet P2 AZ™ directional; BTE = behind the ear; HINT = Hearing in Noise Test; KEMAR = Knowles Electronics Manikin for Acoustic Research; NST = Nonsense Syllable Test; Ρ = Siemens Prisma omnidirectional; PD = Siemens PrismaTM directional; PV = Siemens Prisma omnidirectional with maximum VAD; S = Widex Senso™ omnidirectional; SD = Widex Senso directional; SNR = signal-to-noise ratio; TD = threshold of discomfort; VAD = voice activity detector; WDRC = wide dynamic range compression Article published online: 28 April 2022 © 1999. American Academy of Audiology. This article is published by Thieme. Thieme Medical Publishers, Inc. 333 Seventh Avenue, 18th Floor, New York, NY 10001, USA