Abstract
Background and Objective The size of laser‐induced coagulated lesions produced in porcine muscle in vitro using a cylindrical diffusing fiber tip and a conductive heat source, made by covering the diffuser with a hollow steel needle, were compared to investigate the influence of charring. Materials and Methods Light from a Nd: YAG laser was utilized for thermotherapy. A theoretical model for calculating tissue temperature was used to predict the experimental results and to simulate in vivo treatments. Results The metal‐covered tip produced carbonization and tissue vaporization that was not found with the diffuser. After 20 min of irradiation at a laser power of 7 W, the coagulated volumes with and without carbonization were found to be 13.1 cm3 (range 12.4–14.1 cm3, n = 4) and 12.2 cm3 (range 11.5–13.4 cm3, n = 4), respectively. Mathematical simulations showed that in unperfused tissue, a diffusing laser heat source produces smaller lesions than does a conductive heat source at the same power, the difference in coagulated volume becoming smaller with increased treatment time and increased power. Conclusion Using cylindrical diffusers, interstitial laser‐induced thermotherapy without carbonization at the fiber tip can be as efficient as treatment with carbonization. Lasers Surg. Med. 22:51–57, 1998.

This publication has 23 references indexed in Scilit: