A 1.83 $\mu$J/Classification, 8-Channel, Patient-Specific Epileptic Seizure Classification SoC Using a Non-Linear Support Vector Machine
Authors:
Open Access
- 13 February 2015
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Circuits and Systems
- Vol. 10 (1) , 49-60
Abstract
A non-linear support vector machine (NLSVM) seizure classification SoC with 8-channel EEG data acquisition and storage for epileptic patients is presented. The proposed SoC is the first work in literature that integrates a feature extraction (FE) engine, patient specific hardware-efficient NLSVM classification engine, 96 KB SRAM for EEG data storage and low-noise, high dynamic range readout circuits. To achieve on-chip integration of the NLSVM classification engine with minimum area and energy consumption, the FE engine utilizes time division multiplexing (TDM)-BPF architecture. The implemented log-linear Gaussian basis function (LL-GBF) NLSVM classifier exploits the linearization to achieve energy consumption of 0.39 μ J/operation and reduces the area by 28.2% compared to conventional GBF implementation. The readout circuits incorporate a chopper-stabilized DC servo loop to minimize the noise level elevation and achieve noise RTI of 0.81 μ Vrms for 0.5-100 Hz bandwidth with an NEF of 4.0. The 5 × 5 mm 2 SoC is implemented in a 0.18 μm 1P6M CMOS process consuming 1.83 μ J/classification for 8-channel operation. SoC verification has been done with the Children's Hospital Boston-MIT EEG database, as well as with a specific rapid eye-blink pattern detection test, which results in an average detection rate, average false alarm rate and latency of 95.1%, 0.94% (0.27 false alarms/hour) and 2 s, respectively.Keywords
This publication has 39 references indexed in Scilit:
- Closed-Loop Control of Epilepsy by Transcranial Electrical StimulationScience, 2012
- EEG-based neonatal seizure detection with Support Vector MachinesClinical Neurophysiology, 2011
- A 5.2 mW Self-Configured Wearable Body Sensor Network Controller and a 12 $\mu$W Wirelessly Powered Sensor for a Continuous Health Monitoring SystemIEEE Journal of Solid-State Circuits, 2009
- Economic impact of epilepsy in the United StatesEpilepsia, 2009
- NON-INVASIVE COMPUTERIZED SYSTEM FOR AUTOMATICALLY INITIATING VAGUS NERVE STIMULATION FOLLOWING PATIENT-SPECIFIC DETECTION OF SEIZURES OR EPILEPTIFORM DISCHARGESInternational Journal of Neural Systems, 2009
- Cardiac arrest predictability in seizure patients based on emergency medical dispatcher identification of previous seizure or epilepsy historyResuscitation, 2007
- Kerneltron: support vector "machine" in siliconIEEE Transactions on Neural Networks, 2003
- Severe Myoclonic Epilepsy of Infancy: Extended Spectrum of GEFS+?Epilepsia, 2001
- PhysioBank, PhysioToolkit, and PhysioNetCirculation, 2000
- Seizure Disorders: The Changes With AgeEpilepsia, 1992