Determination of Metaldehyde in Suspected Cases of Animal Poisoning Using Gas Chromatography−Ion Trap Mass Spectrometry

Authors:

Abstract
A method was developed to detect the molluscicide metaldehyde in samples of stomach contents for forensic toxicology investigations. Gas chromatography−ion trap mass spectrometry in full-scan mode was used to identify and quantify metaldehyde. The limit of detection based on mass chromatograms for the m/z 89 ion was 3 μg/g. Mean recoveries from six different spiked samples were 74% at 25 μg/g and 94% at 500 μg/g. The relative standard deviation of six replicate determinations of a sample containing 632 μg/g metaldehyde was 7.3%. Keywords: Metaldehyde; molluscicide; stomach contents; toxicology; GC−MS

This publication has 4 references indexed in Scilit: