Forehead Reflectance Oximetry: A Clinical Comparison with Conventional Digit Sensors during Laparotomic and Laparoscopic Abdominal Surgery
- 16 August 2007
- journal article
- research article
- Published by Springer Nature in Journal of Clinical Monitoring and Computing
- Vol. 21 (5) , 271-276
Abstract
This study compared the performance of forehead reflectance and conventional pulse oximetry (SpO2) in mechanically ventilated, anesthetized patients undergoing major abdominal surgery with either laparoscopic or laparotomic techniques. SpO2 was continuously measured both with a forehead reflectance and a conventional digit sensor in 20 ASA physical status I–III, anesthetized patients undergoing either a laparotomic (group Laparotomy, n = 10) or laparoscopic (group Laparoscopy, n = 10) major abdominal surgery. SpO2 values measured with the two sensors were continuously recorded at 10-second intervals during the entire procedure, and then analyzed for consistency. In group Laparotomy finger and forehead readings remained consistently similar during the study period; on the contrary, in group Laparoscopy forehead SpO2 readings showed a much wider variability. The mean (95% Confidence Intervals) difference between finger and forehead SpO2 readings was 2.0% (−1.3% to +6.0%) in group Laparotomy and 2.5% (−3.5% to +8.4%) in group Laparoscopy (p = 0.001); however, in group Laparoscopy the mean difference between digit and forehead SpO2 values increased from 0.2% (CI95: −2.1% to +2.5%) during the first hour to 4.5% (CI95: −2.5% to +11–5%) in the second hour (p = 0.0005), and 3.1% (CI95: −2.4% to + 8.6%) in the third hour of surgery (p = 0.0005). Clinically relevant desaturation (decrease of SpO2 < 89% for ≥30 s) detected with the forehead sensor in the Laparoscopy group was significantly more frequent and longer lasting than with conventional digit sensor. No differences were observed in group Laparotomy. Forehead reflectance oximetry is as accurate as conventional digit based oximetry in mechanically ventilated patients undergoing laparotomic surgery in the supine position, but is significantly influenced by patient positioning and pneumoperitoneum during laparoscopic surgery.Keywords
This publication has 19 references indexed in Scilit:
- Forehead Pulse OximetryAnesthesiology, 2006
- The Effect of Venous Pulsation on the Forehead Pulse Oximeter Wave Form as a Possible Source of Error in Spo2 CalculationAnesthesia & Analgesia, 2005
- The desaturation response time of finger pulse oximeters during mild hypothermia*Anaesthesia, 2004
- Forehead oximetry in critically ill patients: the case for a new monitoring siteRespiratory Care Clinics of North America, 2004
- Accuracy of pulse oximetry readings in an animal model of low perfusion caused by emerging pneumonia and sepsisIntensive Care Medicine, 2004
- Pulse oximetry during low perfusion caused by emerging pneumonia and sepsis in rabbitsCritical Care Medicine, 2002
- Accuracy of pulse oximetry in the intensive care unitIntensive Care Medicine, 2001
- The Effects of Motion on the Performance of Pulse Oximeters in Volunteers (Revised publication)Anesthesiology, 1997
- Limitations of forehead pulse oximetryJournal of Clinical Monitoring and Computing, 1995
- NON-INVASIVE MONITORING OF FINGER ARTERIAL PRESSURE IN PATIENTS WITH RAYNAUD'S PHENOMENON: EFFECTS OF EXPOSURE TO COLDBritish Journal of Anaesthesia, 1990