Controlling radiofrequency-induced currents in guidewires using parallel transmit
- 17 December 2014
- journal article
- research article
- Published by Wiley in Magnetic Resonance in Medicine
- Vol. 74 (6) , 1790-1802
Abstract
Purpose Elongated conductors, such as pacemaker leads, neurostimulator leads, and conductive guidewires used for interventional procedures can couple to the MRI radiofrequency (RF) transmit field, potentially causing dangerous tissue heating. The purpose of this study was to demonstrate the feasibility of using parallel transmit to control induced RF currents in elongated conductors, thereby reducing the RF heating hazard. Methods Phantom experiments were performed on a four‐channel parallel transmit system at 1.5T. Parallel transmit “null mode” excitations that induce minimal wire current were designed using coupling measurements derived from axial B1+ maps. The resulting current reduction performance was evaluated with B1+ maps, current sensor measurements, and fluoroptic temperature probe measurements. Results Null mode excitations reduced the maximum coupling mode current by factors ranging from 2 to 80. For the straight wire experiment, a current null imposed at a single wire location was sufficient to reduce tip heating below detectable levels. For longer insertion lengths and a curved geometry, imposing current nulls at two wire locations resulted in more distributed current reduction along the wire length. Conclusion Parallel transmit can be used to create excitations that induce minimal RF current in elongated conductors, thereby decreasing the RF heating risk, while still allowing visualization of the surrounding volume. Magn Reson Med 74:1790–1802, 2015.Keywords
Funding Information
- National Institutes of Health (R01-EB008108, R33-CA118276, P01-CA159992, T32-HL007846)
- GE Healthcare
This publication has 35 references indexed in Scilit:
- Reduction of the radiofrequency heating of metallic devices using a dual‐drive birdcage coilMagnetic Resonance in Medicine, 2012
- Ensuring safety of implanted devices under MRI using reversed RF polarizationMagnetic Resonance in Medicine, 2010
- B1 mapping by Bloch‐Siegert shiftMagnetic Resonance in Medicine, 2010
- Measuring RF‐induced currents inside implants: Impact of device configuration on MRI safety of cardiac pacemaker leadsMagnetic Resonance in Medicine, 2009
- Magnitude least squares optimization for parallel radio frequency excitation design demonstrated at 7 Tesla with eight channelsMagnetic Resonance in Medicine, 2008
- Interventional cardiovascular magnetic resonance: still tantalizingJournal of Cardiovascular Magnetic Resonance, 2008
- Reduction of transmitter B1 inhomogeneity with transmit SENSE slice‐select pulsesMagnetic Resonance in Medicine, 2007
- Parallel RF transmission with eight channels at 3 TeslaMagnetic Resonance in Medicine, 2006
- Parallel excitation with an array of transmit coilsMagnetic Resonance in Medicine, 2004
- Effects of magnetic resonance imaging on cardiac pacemakers and electrodesAmerican Heart Journal, 1997