Absorbed dose water calorimeter
Author:
- Steve R. DomenSDSteve R. Domen
- 1 March 1980
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 7 (2) , 157-159
Abstract
Advantage was taken of the low thermal diffusivity of water and the imperviousness of polyethylene film to water to construct a calorimeter for directly measuring absorbed dose in that medium. An ultrasmall bead thermistor was sandwiched between two thin films stretched on polystyrene rings and immersed in an unregulated water bath. Ten cobalt‐60 irradiation runs were made with a precision of 0.5% mean error of the mean at a dose rate of 66 mGy/s. Further development is directed toward a standard instrument that can be used in a medical therapy beam.Keywords
Funding Information
- National Cancer Institute
This publication has 1 reference indexed in Scilit:
- Measurements of heat capacity and heat of vaporization of water in the range 0 degrees to 100 degrees CJournal of Research of the National Bureau of Standards, 1939