Computational Study on the Effects of Valve Orientation on the Hemodynamics and Leaflet Dynamics of Bioprosthetic Pulmonary Valves
- 6 September 2024
- journal article
- research article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 146 (12) , 1-44
Abstract
Pulmonary valves do not display a fibrous annulus as do other valves in the heart; thus, pulmonary valves can be implanted at multiple orientations and locations within the right ventricular outflow tract (RVOT). This gives surgeons more freedom when implanting the valve but it also results in uncertainties regarding placement, particularly with respect to valve orientation. We investigate the pulmonary artery hemodynamics and valve leaflet dynamics of pulmonary valve replacements (PVRs) with various orientations via fluid–structure interaction (FSI) models. A canonical model of the branching pulmonary artery is coupled with a dynamic model of a pulmonary valve, and from this we quantify the effect of valve implant orientation on the postvalvular hemodynamics and leaflet dynamics. Metrics such as turbulent kinetic energy (TKE), branch pulmonary artery flow distributions, projected valve opening area (PVOA), and pressure differentials across the valve leaflets are analyzed. Our results indicate that off-axis orientation results in higher pressure forces and flow and energy asymmetry, which potentially have implications for long-term durability of implanted bioprosthetic valves.Keywords
This publication has 17 references indexed in Scilit:
- A computational study of the hemodynamics of bioprosthetic aortic valves with reduced leaflet motionJournal of Biomechanics, 2021
- In Vitro Assessment of Right Ventricular Outflow Tract Anatomy and Valve Orientation Effects on Bioprosthetic Pulmonary Valve HemodynamicsCardiovascular Engineering and Technology, 2021
- Flow physics of normal and abnormal bioprosthetic aortic valvesInternational Journal of Heat and Fluid Flow, 2020
- Trends and outcomes of pulmonary valve replacement in tetralogy of FallotInternational Journal of Cardiology, 2020
- Implantation Depth and Rotational Orientation Effect on Valve-in-Valve Hemodynamics and Sinus FlowThe Annals of Thoracic Surgery, 2018
- A moving-least-squares immersed boundary method for simulating the fluid–structure interaction of elastic bodies with arbitrary thicknessJournal of Computational Physics, 2016
- Blood Flow Simulations for the Design of Stented Valve Reducer in Enlarged Ventricular Outflow TractsCardiovascular Engineering and Technology, 2015
- Valve-in-Valve Hemodynamics of 20-mm Transcatheter Aortic Valves in Small BioprosthesesThe Annals of Thoracic Surgery, 2011
- Geometry and dimensions of the pulmonary artery bifurcation in children and adolescents: assessment in vivo by contrast-enhanced MR-angiographyThe International Journal of Cardiovascular Imaging, 2010
- A versatile sharp interface immersed boundary method for incompressible flows with complex boundariesJournal of Computational Physics, 2008