Agreement of ocular biometric measurements in young healthy eyes between IOLMaster 700 and OA-2000
Open Access
- 21 February 2020
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 10 (1) , 1-6
Abstract
This prospective cross-sectional study aimed to evaluate the agreement of two new biometers for measuring ocular biometric parameters in young healthy eyes. Ocular biometric parameters were measured using IOLMaster 700 and OA-2000. Power vector analyses of Cartesian (J0) and oblique (J45) components of corneal astigmatism were performed. The right eyes of 103 healthy volunteers were analyzed. The 95% limits of agreement ranged from −0.03 to 0.03 mm, −0.08 to 0.07 mm, −0.18 to 0.18 diopters (D), −1.09 to 1.16 D, −1.18 to 1.15 D for axial length (AL), anterior chamber depth (ACD), mean keratometry, J0 and J45 respectively, which were all comparable between the two biometers, while significant differences were detected in lens thickness (LT), central corneal thickness (CCT), white-to-white (WTW) and pupil diameter (PD). Predicted intraocular lens (IOL) powers were comparable between the two biometers by Haigis and Barrett Universal II formulas, while not by SRK/T, Hoffer Q and Holladay 2. Excepting CCT, WTW and PD meaurements, IOLMaster 700 and OA-2000 have excellent agreement on ocular biometric measurements and astigmatism power vectors, which provides more options for ocular biometric measurements and enables constant optimization for IOL power calculation.Keywords
Funding Information
- Health and Family Planning Commission of Sichuan Province (18ZD022, 18ZD022, 18ZD022, 18ZD022, 18ZD022)
This publication has 30 references indexed in Scilit:
- Biometry measurements using a new large-coherence–length swept-source optical coherence tomographerJournal of Cataract & Refractive Surgery, 2016
- Accuracy of Intraocular Lens Power Formulas Involving 148 Eyes with Long Axial Lengths: A Retrospective Chart-Review StudyJournal of Ophthalmology, 2015
- Comparison of 3 biometry devices in cataract patientsJournal of Cataract & Refractive Surgery, 2015
- Clinical comparison of a new swept-source optical coherence tomography–based optical biometer and a time-domain optical coherence tomography–based optical biometerJournal of Cataract & Refractive Surgery, 2015
- Comparison of ocular biometry and intraocular lens power using a new biometer and a standard biometerJournal of Cataract & Refractive Surgery, 2014
- Statistical methods for conducting agreement (comparison of clinical tests) and precision (repeatability or reproducibility) studies in optometry and ophthalmologyOphthalmic and Physiological Optics, 2011
- Calculation of intraocular lens power: a reviewActa Ophthalmologica Scandinavica, 2007
- Effect of Central Corneal Thickness, Corneal Curvature, and Axial Length on Applanation TonometryArchives of Ophthalmology (1950), 2006
- Sources of error in intraocular lens power calculationJournal of Cataract & Refractive Surgery, 1992
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986